Atec Panametrics-Olympus-IPLEX-FX-LX User Manual

Iplex fx/lx stereo measurement reference guide, A distance, B point-to-line c offset (fx only) d depth

Advertising
background image

1

6

2

3

4

5

Point
Indicates a specified measurement

point or reference point.

Cursor
Specifies a

measurement point

or reference point.

Instruction message box
Shows instructions, conditions, etc.

Displaying the measurement method

Corresponding point

Left image area

Right image area

Object distance
Displays object distance at cursor

position.

<For LX>

• Attach a stereo optical adapter to the distal end of a

scope while the live screen is displayed.

<For FX/LX 8.5-mm type>

Select the stereo optical adapter to be used in the

adapter selection screen, and press the [MEAS/ENTER]

joystick.

<For LX 4-mm and 6-mm types>

Confirm the name of the stereo optical adapter and the

set number, and then select "OK". Press the

[MEAS/ENTER] joystick.

• When using a stereo optical adapter for the first time,

installation will be required. Select the [NEW STEREO

ADAPTER] option from the menu and follow the

on-screen directions. For full details please refer to the

instruction manual.

• Immediately after executing stereo measurement or

selecting "METHOD" in the stereo measurement

screen, the measurement method (measurement mode)

selection screen is displayed.

• Refer to the following page for the usable measurement

method (measurement mode). (Some functions are not

available for LX.)

• Tilt the [ZOOM] lever toward [W] to activate the

measuring distance to object function.

• Highest accuracy is achieved when green zone is

displayed in object distance. To achieve highest

measurement accuracy, bring the insertion tube to an

object as close as possible.

• Push the

[MEAS/ENTER]

joystick to proceed

to the stereo

measurement

screen.

• Object distance is

ceaselessly

displayed in the

stereo

measurement

screen.

Object distance

Measurement

point cursor

Industrial Videoscope

PREPARATIONS FOR STEREO

MEASUREMENT

SPOT RANGING —

OBJECT DISTANCE MEASUREMENT

MEASUREMENT METHOD (MEASUREMENT MODE)

A DISTANCE

Distance mode is for measuring the line segment between

two measurement points
1. Place the cursor on one end of the line segment and press

MEAS/ENTER on the joystick to specify point A.

2. Place the cursor on the other end of the line segment and press

MEAS/ENTER on the joystick to specify point B.

l

The length of line segment A-B is displayed.

Point-to-Line and offset mode are for measuring the distance

from a measurement point to the reference line defined by

two points. In Offset mode, an offset line will additionally be

displayed from the measurement point, in parallel to the

reference line.
1. Place the cursor on both ends of the line segment and press

MEAS/ENTER on the joystick to specify points A and B.

l

Reference line A-B is shown.

2. Place the cursor on the point from which you want to measure the

distance to reference line A-B and press MEAS/ENTER on the

joystick to specify point C.

l

The distance between reference line A-B and point C (or Offset

Line) is displayed. (Offset mode is only available for IPLEX FX

models.)

B POINT-TO-LINE

C OFFSET (FX only)

D DEPTH

Depth mode is for measuring the distance from a

measurement point to the reference plane defined by three

reference points.
1. Place the cursor on points A, B and C to define the reference

plane, then press MEAS/ENTER on the joystick to specify points

A, B and C.

l

Reference plane ABC is shown.

2. Place the cursor on the point from which you want to measure the

distance to reference plane ABC and press MEAS/ENTER on the

joystick to specify measurement point D.

l

The distance between reference plane ABC and point D is

displayed. A positive value (height) means the point is closer to

the tip of the scope than the reference plane, and a negative

value (depth) means the point is farther away.

E AREA/LINES (FX only)

Area mode is for measuring the area defined by multiple

measurement points. Lines mode is for measuring the total

length of lines defined by multiple measurement points.
1.

Place the cursor on points to define the area/lines and press

MEAS/ENTER on the joystick to specify the measurement points.

l

The total length of lines defined by multiple measurement points is

displayed at the bottom of the screen.

Note:

Up to 20 measurement points can be specified per measurement

session.

2.

When measuring an area, one additional point must be positioned to

define the complete subject outline. This point (point 6 in the example

above) needs to be located where the line between this point and the

last point (point 5) cross across the first drawn line.

l

The area enclosed by lines is displayed at the bottom of the screen.

F PROFILE (FX only)

Profile mode is for displaying the surface profile of a line

connecting two points.
1. Place the cursor on both ends of the

line whose surface profile you want to

inspect and press MEAS/ENTER on

the joystick to specify the two points.

l

The surface profile obtained by

sectioning a target object with the plane

defined by the line connecting two

specified points and the optical center

is displayed. The surface profile

graphics overlaid on the right screen

show the profile viewed from the

direction perpendicular to the optic

axis.

G MULTI (FX only)

Multi mode automatically detects and displays the

measurement values of a region specified by two

reference points. The width, depth and surface area

of any missing material will be measured.

1. Place the cursor on each side of the material loss and press

[MEAS/ENTER] on the joystick to specify the reference points.

The region of missing material will be drawn in the left and right

screen. The width, depth and area dimension of missing material

will be displayed at the bottom of screen.

SAVING YOUR STEREO

MEASUREMENT RESULT

Press and hold the [FRZ/REC] button for at least two seconds to

save your stereo measurement image with the measurement

result.

"STILL" indicator will confirm that the image was recorded.

Press [VIEW] button to display the saved image at any time.

STEREO MEASUREMENT

ADVANCED FUNCTIONS

Refer to the Instructions for other detailed stereo

measurement functions.
CLEAR/ALL CLEAR - Delete one or all measurement

point (s).

UNIT - Choose the unit of measurement (millimeters or

inches).

MODIFY (FX only) - Fine adjustments to cursor

placement on left and right images.

TITLE (FX only) - Enter a title in the measurement

screen center (up to 30 alphanumerics).

CURSOR (FX only) - Choose the color, shape and

speed of cursor.

ZOOM WINDOW (FX only) - Activated by moving the

[ZOOM] lever towards the [T] position to provide 2x to

4x magnification.

SUBPIXEL POINTING (FX only) - When the zoom ratio

is greater than 2x, fine mode will be activated and you will

have better fine cursor control under this mode. "F" is

highlighted in the central cursor zoom window while fine

mode is active.

REMEASURMENT OF SAVED IMAGES

Remeasurement can be done via the IPLEX FX, IPLEX

LX system or IPLEX VIEWER PLUS software.

MEASUREMENT SCREEN OVERVIEW

BASIC FUNCTIONS AND TERMS

Point

Indicates a specified measurement

point or reference point.

Cursor zoom window

Displays zoomed image around a specified

point.

Cursor

Specifies a measurement point

or reference point.

Instruction message box

Shows instructions, conditions, etc.

Measurement mode indicator

Corresponding point

Input

region in

left image

Input region in

right image

Corresponding point

zoom window

Displays zoomed image around

corresponding point.

Object distance

Displays object distance at cursor position.

Stereo measurement

message box

Shows measurement results, etc.

Zoom magnification

indicator

Indicates magnification.

[BRT] lever
Adjusting the

brightness

[MENU/EXIT] button
Menu display

[ANGLE] joystick
Angulation control
[LIVE/GAIN] button
Live image display/gain
[MEAS/ENTER] joystick
Menu selection/measurement
[VIEW] button
Saved Image Display

Press once to display last recorded

image or hold for thumbnail selection.
Grip

[ZOOM] lever
Zooming the image

[FRZ/REC] button
Freeze/record observed image

(two locations: left and right)

Hanger
To hang the control unit on the

main unit.
[ANGLE LOCK] lever
To lock the angulation operation

Joystick guard

OLYMPUS CORPORATION obtains the ISO9001/ISO14001.

Specifications, design and accessories are subject to change without any notice or obligation on the part of the manufacturer.

OLYMPUS CORPORATION

Shinjuku Monolith, 3-1 Nishi-Shinjuku 2-chome, Shinjuku-ku, Tokyo 163-0914, Japan

KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT) LTD.

KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom

OLYMPUS NDT INC.

48 Woerd Ave Waltham, MA 02453, U.S.A.

OLYMPUS SINGAPORE PTE LTD.

491B River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373

Notes: For proper and safe operation, consult the manual before use.

Warning: Never use this instrument to observe human or animal body cavities. This could cause injury.

Never use this instrument in an area exposed to flammable gas. This could cause a fire or

explosion.

Never use this instrument in an environment filled with fine particles such as metal dust.

Measurement

point

Measurement

point

Distance

Measurement

point

Reference

point

Perpendicular

distance

Reference

line

Offset Line

(Offset Only)

Reference

point

Reference

point

Reference

point

Measurement

point

Reference

surface

Depth or height

A

D

C

B

Reference point

Reference point

Area

Depth

Width

OLYMPUS EUROPA HOLDING GMBH

Wendenstrasse 14-18, 20097 Hamburg, Germany

OLYMPUS MOSCOW LIMITED LIABILITY COMPANY

House 27, Bld.8, Electrozavodskaya Str Moscow, 107023 Russian Federation

OLYMPUS AUSTRALIA PTY LTD.

31 Gilby Road, Mount Waverley, Victoria 3149, Australia

A

B

C

A

B

Measurement

point

Measurement

point

Measurement

point

Measurement

point

MM0232 04

IPLEX FX/LX STEREO

MEASUREMENT

REFERENCE GUIDE

<For FX>

Advertising