Version 07.00, August 2013

20

Rohde & Schwarz R&S

®

ZNB Vector Network Analyzer

Noise level

at 1 kHz measurement bandwidth,

normalized to 1 Hz

R&S

®

ZNB4 and R&S

®

ZNB8

9 kHz to 50 kHz

< –115 dBm (1 Hz)

50 kHz to 50 MHz

< –120 dBm (1 Hz)

50 MHz to 4 GHz

< –130 dBm (1 Hz)

4 GHz to 6.5 GHz

< –125 dBm (1 Hz)

6.5 GHz to 8.5 GHz

< –120 dBm (1 Hz)

R&S

®

ZNB20

100 kHz to 300 kHz

< –105 dBm (1 Hz)

300 kHz to 1 MHz

< –110 dBm (1 Hz)

1 MHz to 10 MHz

< –115 dBm (1 Hz)

10MHz to 100 MHz

< –120 dBm (1 Hz)

100MHz to 10 GHz

< –125 dBm (1 Hz)

10 GHz to 20 GHz

< –120 dBm (1 Hz)

R&S

®

ZNB40

10 MHz to 50 MHz

< –100 dBm (1 Hz)

50 MHz to 100 MHz

< –110 dBm (1 Hz)

100 MHz to 500 MHz

< –115 dBm (1 Hz)

500 MHz to 20 GHz

< –125 dBm (1 Hz)

20 GHz to 40 GHz

< –115 dBm (1 Hz)

The noise level is defined as the RMS value of the specified noise floor.