ZVA Vector Network Analyzer

DC MEAS 1 V

DC measurement input

Connector type

4-pin mini DIN, female

Voltage range

–1 V to +1 V

Measurement accuracy

2.5 % of reading + 2.5 mV

Resolution

12

bit

Bandwidth

<100

kHz

Input impedance

> 10 kΩ

Damage voltage

30 V

DC MEAS 10 V

DC measurement input

Connector type

4-pin mini DIN, female

Voltage range

–10 V to +10 V

Measurement accuracy

2.5 % of reading + 25 mV

Resolution

12

bit

Bandwidth

<100

kHz

Input impedance

> 10 kΩ

Damage voltage

30 V

PORT BIAS

DC bias input for PORT

Connector type

BNC, female

Maximum nominal input voltage

30 V

Maximum nominal input current

200 mA

Damage voltage

30 V

Damage current

500 mA

MONITOR

IBM-PC-compatible VGA monitor connector, 15-pin D-Sub (for external monitor)

USER CONTROL

several control and trigger signals, 25-pin D-Sub, 3.3 V TTL
for controlling external generators, for limit checks, sweep signals, etc.

FOOT SWITCH 1 and FOOT SWITCH 2

pin 24 and pin 25 (inputs)

control inputs

DRIVE PORT 1 to DRIVE PORT 4

pin 16 to pin 19 (outputs)

indicate driving port

CHANNEL BIT 0 to CHANNEL BIT 3

pin 8 to pin 11 (outputs) channel-specific

user-configurable bits

PASS 1 and PASS 2

pin 13 and pin 14 (outputs)

pass/fail results of limit checks

BUSY

pin 4 (output)

measurements running

READY FOR TRIGGER

pin 6 (output)

ready for trigger

EXT GEN TRIGGER

pin 21 (output)

control signal for external generator

EXT GEN BLANK

pin 22 (input)

handshake signal from external generator

EXTERNAL TRIGGER

pin 2 (input)

trigger input for analyzer

EXT TRIGGER

trigger input for analyzer

Connector type

BNC, female

TTL signal (edge-triggered)

3 V

Polarity (user-selectable)

positive or negative

Minimum pulse width

1 µs

Input impedance

> 10 kΩ