Atec LeCroy-WAVEPRO-950-960 User Manual

Wavepro oscilloscopes, Specifications

Advertising
background image

WavePro Oscilloscopes

Specifications

Vertical System

WavePro 960

DISCONTINUED

WavePro 950

DISCONTINUED

WavePro 940

DISCONTINUED

Input Channels

4

4

4

Analog Bandwidth @ 50
Ohms (-3 dB)

2 GHz*

1 GHz

500 MHz

Bandwidth Limiters

20 MHz, 200 MHz

Input Impedance

50 Ohms ± 1.5%; 10 MOhms // 11 pF typical (using PP005 probe)

Input Coupling

1 MOhms : AC, DC, GND; 50 Ohms : DC, GND

Maximum Input

50 Ohms : 5 Vrms; 1 MOhms : 100 Vmax (peak AC =5 kHz + DC)

Vertical Resolution

8 bits; up to 11 bits with enhanced resolution (ERES)

Sensitivity

50 Ohms : 1 mV – 1 V/div fully variable ; 1 MOhms : 1 mV – 2 V/div fully variable

DC Accuracy

± 2.0% full scale + 1.5% offset value @ gain > 10 mV

Offset Accuracy

± (1.5% + 0.5% of full scale + 1 mV)

Offset Range

50 Ohms or 1 MOhms : 1 mV – 4.99 mV/div: ±400 mV

50 Ohms : 5 mV – 99 mV/div: ±1 V; 0.1 V – 1 V/div: ±10 V

1 MOhms: 5 mV – 100 mV/div: ±1 V; 101 mV – 2 V/div: ±20 V

Isolation — Channel-to-
Channel

> 250:1 at same V/div settings

Timebase System

Timebases

Main and up to four independent zoom traces simultaneously

Ranges

200 ps/div – 1000 s/div

Clock Accuracy

=10 ppm

Interpolator Resolution

5 ps

External Clock Frequency

500 MHz maximum, 50 Ohms, or 1 MOhms impedance

Roll Mode – Operating Range

time/div 500 ms – 1000 s/div or sample rate < 100 kS/s max

External Reference

10 MHz timebase reference clock available with input on rear panel

External Timebase Clock

500 MHz maximum external sample clock input on front panel EXT BNC

Acquisition System

Single-Shot Sample Rate

1 Channel Max.

16 GS/s

16 GS/s

8 GS/s

2 Channels Max.

8 GS/s

8 GS/s

8 GS/s

3 – 4 Channels Max.

4 GS/s

4 GS/s

4 GS/s

Advertising