Atec Quadtech-G32 User Manual

G u a rd i a n, For electronic medical devices

Advertising
background image

G u a rd i a n

6100 Production Safety A n a ly ze r

USES:

• Production Testing of Medical

Electronic Products and Instruments
in Accordance with UL, CSA, and IEC
standards such as UL1950, IEC601-1,
UL2601-1, UL1563 and UL544

• Electrical Safety Compliance Testing

on a Variety of Products and Devices

FEATURES:

• Programmable output voltage to 5KV

AC and 6KV DC

• Ground Bond Testing to 30A AC with

Adjustable Limit

• Line/Earth Leakage Current

Measurement(UL2601-1/UL1950)with
5 Human Body Models

• Insulation Resistance Measurements

from 100k

to 50G

• Monitoring of DUT Current Draw

• Programmable Ramp and Test Times

• Storage of 50 Tests Setups with 10

Steps per Setup

• Continuous Leakage Current

Monitoring

• Trip Current Programmable to 40mA

AC and 20mA DC

• Front Panel Lockout via Password

• Standard IEEE and Remote Control

Interfaces

• Optional RS232 Interface

Introduction

The Guardian 6100 is designed for fast, easy, production testing of single phase
electronic medical devices and instruments. Without changing the connection to the
device, the tester is able to perform five essential electrical safety tests in one unit,
including AC Hipot, DC Hipot, Insulation Resistance, Line Leakage Current and Ground
Bond measurements. Performing all these electrical safety tests in one box increases
production by reducing test time and eliminating multiple test stations.

Description

AC Hipot Test

Performs AC dielectric testing (hipot) over the voltage range from

50 to 5000VAC RMS. Leakage current detection is programmable from 1µA to 40mA,
making the 6100 ideal for testing medical products and instruments having a wide range
of leakage currents.

DC Hipot Te s t

DC dielectric testing is possible from 50 to 6000VDC with leakage

current detection down to 0.1µA. The maximum total output current of 20mA allows
quick charging of capacitive devices and products.

Insulation Resistance

The insulation resistance test (similar to DC Hipot)

calculates and displays a product’s insulation resistance value in ohms. This resistance
can be measured over the range of 100k

to 50G

with test voltages programmable

from 50 to 1000VDC.

Ground Bond

The ground bond test, sometimes referred to as a high current

continuity test, can be programmed from 1 to 30A for verifying the integrity of a product’s
ground system. Resistance measurements are displayed and a high limit programmable
between 0.1m

and 510m

.

Line/Earth Leakage Current

A line/earth leakage test can be performed directly

on the 6100 in four possible configurations; normal operating conditions, reverse line,
or a single fault normal and single fault reverse mode. Five different human body models
can be selected in accordance with UL, IEC, and other standards. Leakage current limits
are programmable from 0.1µA to 9.999mA.

Note: This test requires input isolation transformer for proper operation

1-800-253-1230

Fax 1-508-485-0295

Intl. 1-508-229-0806

For more information about special purchase,

rent & lease options, call

For more detailed specifications, visit

www.quadtech.com

For Electronic Medical Devices

HIPOT • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS • DECADES • STANDARDS

Advertising