Magnetic-Field Test System / Low-Frequency Test System
for Emission and Immunity Tests / MTS-800

Technical specifications

Voltage input (Analyzer)

Frequency range

DC - 250 kHz

Input impedance

1 M

Ω

/ 50

Ω

switchable

Connector

XLR, unbalanced

Max. input voltage

100 V continuous (attenuator autoset at overvoltage); 10 V at 50

Ω

Gain

-20/0/20 dB Preamplifier, 0/20/40 dB ADC Amplifier; Self-calibration with ultra stable on-board reference

Current input

Frequency range

DC - 250 kHz

Shunts

10 m

Ω

/ 1

Ω

/ 100

Ω

Max. input current

20 A continuous (overload protection); 1

Ω

and 100

Ω

shunt are protected by an additional 1.5 A fuse

Connector

4 mm safety jack (+, -) measurement via insulation amplifier or input jacks

Measurement range

20 A, 10 A, 1 A, 100 mA, 10 mA, 1 mA automatic offset and gain;
Self-calibration with ultra stable on-board reference

AD converter

Resolution

16 Bit

Sampling rate

1.25 MSPS

Aliasingfilter

0.01 dB Tschebyscheff filter, fg = 260 kHz; filter may be switched off

Generator

Frequency range

DC - 250 kHz

Output impedance

50

Ω

Connector

BNC, unbalanced

Signal

Sine wave / triangular /square wave/ DC

Amplitude

0 to 10 VAC, -10 V to +10 VDC

Resolution

12 Bit (2.5 mV), Switchable - 20 dB Attenuator; Self-calibration with ultra stable on-board reference

Amplifier

Frequency range

DC - 1 MHz

Connector

4 mm safety jacks (output); BNC, unbalanced (input)

Current

16 Arms

Voltage

50 Vrms / 75 VDC

Distortion (DC-100 kHz, load

≥ 4

Ω

)

< 0.10 %

General data

EUT control / Connector

9-pin Sub-D; RS232

Connection to Computer

USB

Temperature range

0 to 40 °C

Warm-up time

15 min.

Housing

19”-Subrack or desktop case

Dimensions (W x H x D)

449 x 177 x 580 mm

Weight (shipping)

approx. 40 kg (net 34 kg)

Gain

10 ± 0.1 % (±0.01 % / °C)

PAGE // 24