Millivolt measurement and source, Temperature, thermocouples – Atec Fluke-724 User Manual

Page 51

Advertising
background image

Temperature Calibrator

Specifications

41

Millivolt Measurement and Source*

Range

Resolution

Accuracy

-10 mV to 75 mV

0.01 mV

±

(0.025 % + 1 count)

Maximum input voltage: 30 V

Temperature coefficient -10

°

C to 18

°

C, +28

°

C to

55

°

C:

±

0.005 % of range per

°

C

*Select this function by pressing

T

. The signal is

available at the thermocouple miniplug connector.

Temperature, Thermocouples

Type

Range

Measure and Source

Accuracies (ITS-90)

J

-200 to 0

°

C

0 to 1200

°

C

1.0

°

C

0.7

°

C

K

-200 to 0

°

C

0 to 1370

°

C

1.2

°

C

0.8

°

C

T

-200 to 0

°

C

0 to 400

°

C

1.2

°

C

0.8

°

C

E

-200 to 0

°

C

0 to 950

°

C

0.9

°

C

0.7

°

C

R

-20 to 0

°

C

0 to 500

°

C

500 to 1750

°

C

2.5

°

C

1.8

°

C

1.4

°

C

S

-20 to 0

°

C

0 to 500

°

C

500 to 1750

°

C

2.5

°

C

1.8

°

C

1.5

°

C

B

600 to 800

°

C

800 to 1000

°

C

1000 to 1800

°

C

2.2

°

C

1.8

°

C

1.4

°

C

L

-200 to 0

°

C

0 to 900

°

C

0.85

°

C

0.7

°

C

U

-200 to 0

°

C

0 to 400

°

C

1.1

°

C

0.75

°

C

N

-200 to 0

°

C

0 to 1300

°

C

1.5

°

C

0.9

°

C

Resolution:

J, K, T, E, L, N, U:

0.1

°

C, 0.1

°

F

B, R, S:

1

°

C, 1

°

F