Additional specifications, Frequency specifications – Atec Fluke-5500A User Manual

Page 20

Advertising
background image

5500A Black and White Extended Specifications

Fluke Corporation

19

Additional Specifications

The following paragraphs provide additional specifications for the 5500A Calibrator ac voltage and ac current functions. These
specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5500A has been turned off. All extended range
specifications are based on performing the internal zero-cal function at weekly intervals, or when the ambient temperature changes by more
than 5

°C. (See Chapter 4, Front Panel Operations in the 5500A Operator Manual.)

Frequency Specifications

Frequency

Range

Resolution

1-Year Absolute Uncertainty,

tcal ± 5

°C

Jitter

0.01 - 119.99 Hz

0.01 Hz

25 ppm, ± 1 mHz

2

µs

120.0 - 1199.9 Hz

0.1 Hz

25 ppm, ± 1 mHz

2

µs

1.200 - 11.999 kHz

1.0 Hz

25 ppm, ± 1 mHz

[1]

2

µs

12.00 - 119.99 kHz

10 Hz

25 ppm, ± 15 mHz

140 ns

120.0 - 1199.9 kHz

100 Hz

25 ppm, ± 15 mHz

140 ns

1.200 - 2.000 MHz

1 kHz

25 ppm, ± 15 mHz

140 ns

[1]

± (25 ppm + 15 mHz) above 10 kHz

Advertising