Corrected absolute ac/dc uncertainty, Performance specifications, Fully traceable performance – Atec Fluke-792A User Manual

Page 2: Description, 1 year, t, 12°c, Volt-hertz product: 1x10, Volt-hertz product: 1 x 10

Advertising
background image

Corrected absolute ac/dc uncertainty

1 year, T

CAL

± 12°C

AC/DC uncertainty

±

ppm input

Voltage

Frequency

Voltage

Voltage

Range

Input

10 Hz

20 Hz

40 Hz

100 Hz

1 kHz

10 kHz

20 kHz

50 kHz

100 kHz

300 kHz

500 kHz

800 kHz

1 MHz

22 mV

2 mV

2300

1700

1700

1700

1700

1700

1700

1700

2000

2700

5000

5600

5600

6 mV

780

430

380

380

380

380

380

500

1200

1700

2100

2400

2400

10 mV

500

300

300

300

300

300

300

300

600

800

1400

2000

2000

20 mV

400

200

200

180

180

180

180

250

600

800

1200

1300

1300

220 mV

20 mv

450

250

220

220

220

220

220

280

700

900

1400

1200

1200

60 mV

260

130

90

70

70

70

70

130

270

560

660

650

650

100 mV

240

100

55

55

55

55

55

72

160

300

330

380

380

200 mV

220

90

45

34

34

34

34

72

160

240

270

330

330

700 mV

200 mV

230

90

55

55

55

55

55

83

160

240

270

330

330

600 mV

210

75

35

26

26

26

26

50

65

130

150

170

170

2.2V

600 mV

210

65

30

25

25

25

25

45

60

120

140

140

140

1V

200

60

30

15

15

15

15

40

50

120

130

130

130

2V

200

60

25

10

10

10

10

40

50

120

130

130

130

7V

2V

200

65

30

22

22

22

22

45

55

120

140

160

160

6V

200

60

25

10

10

10

10

40

50

120

130

140

140

22V

6V

200

65

30

22

22

22

22

45

55

120

130

170

170

10V

200

60

30

15

15

15

15

40

50

120

130

150

150

20V

200

60

25

15

15

15

15

40

50

120

130

150

150

70V

20V

200

70

30

25

25

25

25

55

70

130

150

150

150

60V

200

65

25

20

20

20

20

50

65

130

220V

60V

200

70

33

33

33

33

33

65

70

130

100V

200

70

30

20

20

20

20

60

70

200V

200

65

30

18

18

18

18

60

70

1000V

200V

200

95

40

33

33

33

33

65

70

600V

200

95

39

27

27

30

30

60

70

1000V

200

95

30

25

25

27

27

60

70

Volt-hertz product: 1x10

8

@ 100 kHz, 2.2x10

7

@ 1 MHz

Waveform requirements: Sinusoidal, distortion less than 1%

792A AC/DC Transfer Standard Fluke Corporation 4

Corrected ac/dc uncertainties relative to calibration standards

1 year, T

CAL

± 5°C

AC/DC uncertainty

±

ppm input

Voltage

Frequency

Voltage

Voltage

Range

Input

10 Hz

20 Hz

40 Hz

100 Hz

1 kHz

10 kHz

20 kHz

50 kHz

100 kHz

300 kHz

500 kHz

800 kHz

1 MHz

22 mV

2 mV

1300

1200

1100

1100

1100

1100

1100

1300

1500

2000

3000

4000

4000

6 mV

400

325

275

275

275

275

275

400

750

1000

1300

1500

1500

10 mV

300

260

225

225

225

225

225

250

400

650

900

1000

1000

20 mV

250

150

125

125

125

125

125

175

350

500

650

700

700

220 mV

20 mV

300

160

150

130

130

130

130

180

400

550

700

800

800

60 mV

200

50

25

25

25

25

25

50

250

450

450

500

500

100 mV

150

50

12

12

12

12

12

50

150

200

250

300

300

200 mV

80

50

10

10

10

10

10

50

50

200

200

275

275

700 mV

200 mV

70

50

25

25

25

25

25

50

50

200

200

275

275

600 mV

70

19

7

7

7

7

7

30

35

70

125

130

130

2.2V

600 mV

70

19

10

10

10

10

10

30

35

70

125

130

130

1V

70

19

7

7

7

7

7

30

35

70

100

100

100

2V

70

19

7

6

6

7

7

30

35

70

10

100

100

7V

2V

70

19

15

15

15

15

15

30

35

70

125

130

130

6V

70

19

7

6

6

7

7

30

35

70

100

100

100

22V

6V

70

19

15

15

15

15

15

30

35

70

125

130

130

10V

70

19

7

7

7

7

7

30

35

70

100

100

100

20V

70

19

7

7

7

7

7

30

35

70

100

100

100

70V

20V

70

19

15

15

15

15

15

30

35

70

115

115

115

60V

70

19

7

7

7

7

7

30

35

70

220V

60V

70

19

15

15

15

15

15

30

35

100

100V

70

19

7

7

7

7

7

30

50

200V

70

19

7

7

7

7

7

30

50

1000V

200V

70

19

17

16

16

18

18

30

50

600V

100

35

17

16

16

18

18

30

50

1000V

100

35

10

10

10

10

10

30

50

Volt-hertz product: 1 x 10

8

@ 100 kHz, 2.2 x 10

7

@ 1 MHz

Waveform requirements: Sinusoidal, distortion less than 1%.

Corrected absolute ac/dc uncertainty

1 year, T

CAL

± 5°C

AC/DC uncertainty

±

ppm input

Voltage

Frequency

Voltage

Voltage

Range

Input

10 Hz

20 Hz

40 Hz

100 Hz

1 kHz

10 kHz

20 kHz

50 kHz

100 kHz

300 kHz

500 kHz

800 kHz

1 MHz

22 mV

2 mV

2200

1600

1600

1600

1600

1600

1600

1600

1900

2500

4500

5000

5000

6 mV

500

400

350

350

350

350

350

500

1000

1500

1800

1800

1800

10 mV

320

280

280

280

280

280

280

300

500

700

1200

1500

1500

20 mV

280

180

140

125

125

125

125

200

450

650

1000

1000

1000

220 mV

20 mV

350

230

190

190

190

190

190

220

450

650

1000

1100

1100

60 mV

230

120

80

60

60

60

60

120

250

500

600

600

600

100 mV

210

90

50

50

50

50

50

65

150

270

300

350

350

200 mV

200

80

40

30

30

30

30

65

150

220

250

300

300

700 mV

200 mV

220

80

50

50

50

50

50

75

150

220

250

300

300

600 mV

200

70

30

22

22

22

22

45

60

130

140

140

140

2.2V

600 mV

200

65

30

25

25

25

25

45

60

120

140

140

140

1V

190

60

30

15

15

15

15

40

50

115

125

125

125

2V

190

60

25

10

10

10

10

40

50

115

125

125

125

7V

2V

190

65

30

22

22

22

22

45

55

120

140

140

140

6V

190

60

25

10

10

10

10

40

50

120

125

125

125

22V

6V

190

65

30

22

22

22

22

45

55

120

125

140

140

10V

190

60

30

15

15

15

15

40

50

120

125

125

125

20V

190

60

25

15

15

15

15

40

50

120

125

125

125

70V

20V

190

65

30

25

25

25

25

55

70

130

140

140

140

60V

190

60

25

20

20

20

20

50

65

130

220V

60V

190

65

33

33

33

33

33

65

70

130

100V

190

65

30

20

20

20

20

60

70

200V

190

60

27

18

18

18

18

60

70

1000V

200V

190

90

40

33

33

33

33

65

70

600V

190

90

35

27

27

30

30

60

70

1000V

190

90

27

25

25

27

27

60

70

Volt-hertz product: 1x10

8

@ 100 kHz, 2.2x10

7

@1 MHz

Waveform requirements: Sinusoidal, distortion less than 1%

792A AC/DC Transfer Standard Fluke Corporation 3

Performance
specifications

There are three tables of uncer-
tainties included for the 792A.
The first two are absolute and
traceable to the National Insti-
tute of Standards and Technol-
ogy (NIST). Absolute uncertainty
includes time stability, tempera-
ture, line and load regulation,
and the traceability provided by
NIST (see Fluke application note
B0205A, Establishing Traceabil-
ity for a High Performance

AC/DC Transfer Standard). The
third table, Relative Uncertainty,
includes time stability, tempera-
ture, line and load regulation,
but not the traceability to
national or external standards.

These uncertainties apply

when the calibration report,
provided with the 792A, is used
to correct the response of the
transfer standard.

2 Fluke Corporation 792A AC/DC Transfer Standard

The RMS Sensor is designed

to be rugged and reliable. Each
is built to exacting standards by
the Fluke Microelectronics Op-
eration to maintain quality and
consistency, part after part.

Fully traceable
performance

Each 792A is shipped from
Fluke with documented trace-
ability to the US National Insti-
tute of Standards and Technol-
ogy (NIST). Included is a table
of correction factors and uncer-
tainties for measured ac/dc
differences.

To support the traceability

requirements for an instrument
as accurate as the 792A, the
Fluke Metrology Department
developed a unique system to
transfer NIST values to the 792A
production environment. This
system is based on proven tech-
niques developed to maintain a
732A direct voltage standard to
within a few parts in 100 million.

Statistical treatment of data

resulting from successive
intercomparisons of virtually
identical 792As is used to mini-
mize transfer uncertainties. To
achieve even higher perfor-
mance, you can have the trans-
fer uncertainties of your 792A
assigned directly by NIST, or
other national standards organi-
zations. For more information,
request the application note
titled Establishing Traceability
for a High Performance AC/DC
Transfer Standard
.

Periodic calibration and

recertification is available from
Fluke. This service repeats the
original calibration procedure.
A new table of correction factors
and undertainties for measured
ac/dc differences is returned
with your instrument. Order
792A-000.

Description

The 792A consists of four units.
The Transfer Unit is the main
analog component of the 792A
for the 20 mV to 220V ranges.
Stainless Steel Type-N connec-
tors assure low signal loss and
high measurement repeatability.
The separate Power Pack unit
permits the Transfer Unit and
1000V Range Resistor to be
shipped independently for

calibration. The 1000V Range
Resistor isolates the heat gener-
ated at high voltages outside
the main Transfer Unit. The
Range Resistor has a low tem-
perature coefficient, stabilizes
quickly and exhibits small ac/dc
differences. The Transfer Switch
provides for switching between
the ac and dc inputs.

Each 792A is shipped from

the factory traceable to NIST
standards. Each 792A also
includes a table of correction
factors and uncertainties for
measured ac/dc differences.

The 792A consists of four main components: The Transfer Unit, Power Pack, 1000V Range
Resistor and Transfer Switch.

Each 792A includes a table of correction factors and uncertainties for measured ac/dc differences.

Advertising