Measurement function specifications, Sourcing (simulation) function specifications, Confidence interval: k =3 – Atec Fluke-750 Series User Manual

Page 8

Advertising
background image

DC current measurement

Accuracy (% of reading + floor)

Range (full scale)

1 year

2 years

30.000 mA

0.01 % + 5 uA

0.015 % + 7 uA

110.00 mA

0.01 % + 20 uA

0.015 % + 30 uA

Temperature coefficient: (3 % of specified accuracy)/°C from -10 °C to 18 °C

and 28 °C to 50 °C

Normal mode rejection: 90 dB at 50 or 60 Hz nominal and 60 dB at 1200 Hz

and 2200 Hz (HART signals)

Resistance measurement

Accuracy (% of reading + ohms)

Range (full scale)

1 year

2 years

10.000 Ω

0.05 % + 50 mΩ

0.07 % + 70 mΩ

100.00 Ω

0.05 % + 50 mΩ

0.07 % + 70 Ω

1.0000 kΩ

0.05 % + 0.5 Ω

0.07 % + 0.5 Ω

10.000 kΩ

0.1 % + 10 Ω

0.15 % + 15 Ω

Temperature coefficient: (3 % of specified accuracy)/°C from -10 °C to 18 °C

and 28 °C to 50 °C

Maximum input voltage: 50 V dc

Continuity: Continuous tone < 25 Ω, No tone > 400 Ω

Specifications are valid to 110 % of range

Frequency measurement

Resolution Accuracy

Range

1

2 years

1.00 Hz to 110.00 Hz

0.01 Hz

0.05 Hz

110.1 Hz to 1100.0 Hz

0.1 Hz

0.5 Hz

1.101 kHz to 11.000 kHz

0.001 kHz

0.005 kHz

11.01 kHz to 50.00 kHz

0.01 kHz

0.05 kHz

1

For frequencies < 109.99 Hz, specification applies for signals with slew rates > 5 V/ms

Minimum amplitude for Hz measurement: (Squarewaves) 1 Hz to

1 kHz, 300 mV p-p; 1 kHz to 30 kHz, 1.4 V p-p; > 30 kHz, 2.8 V p-p

Maximum input: 1 Hz to 1 kHz, 300 V rms; > 1 kHz, 30 V rms

Input impedance: 4 MΩ

DC voltage measurement

Accuracy (% of reading + floor)

Range (full scale)

1 year

2 years

100.000 mV

0.02 % + 0.005 mV

0.03 % + 0.005 mV

3.00000 V

0.02 % + 0.00005 V

0.03 % + 0.00005 V

30.0000 V

0.02 % + 0.0005 V

0.03 % + 0.0005 V

300.00 V

0.05 % + 0.05 V

0.07 % + 0.05 V

Temperature coefficient: (0.001 % reading + 0.0002 % range)/°C

from -10 °C to 18 °C and 28 °C to 50 °C, 100.000 mV range: 0.001 % of

reading + 0.001 % of range

Input impedance: >4 MΩ

Maximum input voltage: 300 V rms

Normal mode rejection: >100 dB at 50 Hx or 60 Hx nominal

Specifications are valid to 110 % of range (except for 300 V range)

Measurement function specifications

Confidence Interval: k =3

Sourcing (simulation) function specifications

Confidence Interval: k =3

DC voltage output

Accuracy (% of output + floor)

Range (full scale)

1 year

2 years

100.000 mV

0.01 % + 0.005 mV

0.015 % + 0.005 mV

1.00000 V

0.01 % + 0.00005 V

0.015 % + 0.0005 V

15.0000 V

0.01 % + 0.0005 V

0.015 % + 0.0005 V

Temperature coefficient: (0.001 % output + 0.001 % f.s.)/°C

from -10 °C to 18 °C and 28 °C to 50 °C

Maximum output current: 10 mA

Specifications are valid to 110 % of range, 100 mV and 1 V ranges

DC current output

Accuracy (% of output + floor)

Range (full scale)

1 year

2 years

22.000 mA

0.01 % + 0.003 mA

0.02 % + 0.003 mA

Current sink (sim-

0.02 % + 0.007 mA

0.04 % + 0.007 mA

ulate transmitter)

Specification applies from 0.1 mA to 22 mA; below 2 mA typical accuracy is

0.15 % of full scale

Maximum burden voltage: 18 V

Temperature coefficient: 3 % of specified accuracy/°C from -10 °C to 18 °C

and 28 °C to 50 °C

Resistance sourcing

Accuracy (% of output + ohms)

Range

1 year

2 years

10.000 Ω

0.01 % + 10 mΩ

0.015 % + 15 mΩ

100.00 Ω

0.01 % + 20 mΩ

0.015 % + 30 mΩ

1.0000 kΩ

0.02 % + 0.2 Ω

0.03 % + 0.3 Ω

10.000 kΩ

0.02 % + 3 Ω

0.03 % + 5 Ω

Temperature coefficient: 0.01 % f.s./°C from -10 °C to 18 °C and 28 °C to 50 °C

Maximum and minimum current through source resistance:

Maximum Minimum

10 Ω range:

10 mA dc

0.1 mA dc

100 Ω range: 10 mA dc

0.1 mA dc

1.0 kΩ range: 1 mA dc

0.01 mA dc

10 kΩ range: 1 mA dc

0.01 mA dc

Specifications valid to 110 % of range

8 Fluke Corporation 750 Series Brochure

Frequency sourcing

Range

Specification

2 years

Sinewave: 0.1 Hz to 10.99 Hz

0.01 Hz

Squarewave: 0.01 Hz to 10.99 Hz

0.01 Hz

Sine and square 11.00 Hz to 109.99 Hz

0.1 Hz

Sine and square 110.0 Hz to 1099.9 Hz

0.1 Hz

Sine and square 1.100 kHz to 21.999 kHz

0.002 kHz

Sine and square 22.000 kHz to 50.000 kHz

0.005 kHz

Waveform choices: Zero-symmetric sine wave or positive 50 % duty-cycle square wave

Square wave amplitude: 0.1 V to 15 V p-p

Square wave amplitude accuracy:

0.01 kHz to 1 kHz: 1 % p-p output + 75 mV,

1 kHz to 50 kHz: 10 % p-p output + 75 mV

Sine wave amplitude: 0.1 V to 30 V p-p

Sine wave amplitude accuracy, 0.1 Hz to 50 kHz: 3 % p-p output + 75 mV

Maximum input voltage: ± 30 V dc

AC voltage measurement

Range

40 Hz to 500 Hz

Resolution

% of reading + floor

1 year

2 year

3.000 V

0.001 V

.5 % + 0.002 V 1.0 % + 0.004 V

30.00 V

0.01 V

0.5 % + 0.02 V

1.0 % + 0.04 V

300.0 V

0.1 V

0.5 % + 0.2 V

1.0 % + 0.2 V

Input impedance: >4 M Ω and <100 pF

Input coupling: AC

Maximum input voltage: 300 V, IEC 61010 300V CAT II

Temperature coefficient: 5 % of specified accuracy / °C (<18 °C or >28 °C)

Specifications apply for 9 % to 100 % of voltage range.