Atec Fluke-7250 Series User Manual

Series 7250, Ruska pressure controller/calibrator, Technical data

Advertising
background image

Technical Data

Features

Pressure ranges from 0 to 5 and 0 to

3 000 psi (0 to 35 kPa and 0 to 21 MPa)

Model 7250xi and 7250i provide advanced

precision of 0.005 % of reading

Model 7250 provides 0.003 % of full scale

precision

Stability: 0.0075 % of reading per year

Time to setpoint: 20 seconds with no

overshoot

Control stability: 10 ppm

Active matrix color screen with enhanced

navigation menus

Languages: English, French, Chinese,

German, Japanese, Spanish and Italian

Series 7250

Ruska Pressure

Controller/Calibrator

The Series 7250 high speed digital pressure

controllers represent the fifth generation of

automatic pressure controllers from Fluke Cali-

bration and offers unmatched performance, an

active matrix color screen and enhanced control

stability. Utilizing multiple sensors and ranges in

a single instrument, the Series 7250 combines

precision, stability, speed and affordability.

Models 7250xi, 7250i and 7250, up to

2 500 psi (17.2 MPa), use a unique quartz

sensor, the most accurate pressure sensing

technology in a pressure calibrator. Each quartz

sensor is manufactured and tested to provide

the ultimate performance required by a Fluke

Calibration pressure calibrator, ensuring that

every customer receives quality, precision and

stability in their instrument.

Advanced precision

Models 7250xi and 7250i offer advanced per-

cent of reading precision for increased capability

with a single instrument, reducing the invest-

ment required to calibrate a wide variety of

pressure devices and ranges. The Model 7250xi

provides 0.005 % of reading precision from

5 % to 100 % of the instrument’s range. This

unmatched precision is achieved through unique

quartz pressure sensing technology and mul-

tiple quartz sensors in a single instrument. The

7250xi is available with any full scale range

from 20 to 2 500 psi (140 kPa to 17.2 MPa). The

Model 7250i provides 0.005 % of reading from

25 % to 100 % of range and is available with

any full scale range from 5 to 2 500 psi

(35 kPa to 17.2 MPa).

For pressure below the lower threshold of

5% for the 7250xi and 25 % for the 7250i, the

precision becomes 0.005 % of the lower value.

For example, a 1 000 psi (7 MPa) Model 7250xi

provides 0.005 % of reading from 50 to 1 000

psi (35 kPa to 7 MPa); the precision for pressures

from 0 to 50 psi (0 to 35 kPa) is 0.005 % of

50 psi (35 kPa).

Selecting the full scale range of the 7250xi

and 7250i to purchase is simplified with the

percent of reading capability. Simply determine

the highest pressure required and then decide if

0.005 % of reading is required down to 25 % or

5 % to maintain the appropriate calibration ratio.

The complex analysis associated with selecting

the appropriate single or multirange instruments

with percent of full scale performance is no

Advertising