The specifications for the 4100 Series of Ohmmeters are listed in the following paragraphs. In all
cases the specifications are valid for full Kelvin Four-Terminal measurements using connections
having less than 20 milliohms of lead resistance per wire.

Accuracy


The accuracy specifications listed below are valid following a 30 minute warm-up at an ambient
temperature between 15

°C and 35°C for a period of 1 year following calibration, and include the

effects of line voltage variations within the allowed range.




Model 4100ATC

Range Maximum

Input Resolution

Test

Current

Accuracy

.2Ω .19999

Ω 10µΩ

1 A

±0.02% of rdg ±2 digits

2Ω 1.9999

Ω 100µΩ

100 mA

±0.02% of rdg ±2 digits

20Ω 19.999

Ω 1mΩ

10 mA

±0.02% of rdg ±2 digits

200Ω 199.99

Ω 10mΩ

1 mA

±0.02% of rdg ±2 digits

2kΩ 1.9999

kΩ 100mΩ

100 µA

±0.02% of rdg ±2 digits

20kΩ 19.999

kΩ

1Ω

10 µA

±0.02% of rdg ±2 digits

Full Scale Test Voltage: 200mV





Model 4150ATC

Range Maximum

Input Resolution

Test

Current

Accuracy

20mΩ 19.999

mΩ 1µΩ

1 A

±0.03% of rdg ±3 digits

200mΩ 199.99

mΩ 10µΩ

100 mA

±0.03% of rdg ±3 digits

2Ω 1.9999

Ω 100µΩ

10 mA

±0.03% of rdg ±3 digits

20Ω 19.999

Ω 1mΩ

1 mA

±0.03% of rdg ±3 digits

200Ω 199.99

Ω 10mΩ

100 µA

±0.03% of rdg ±3 digits

2KΩ 1.9999

kΩ 100mΩ

10µA

±0.03% of rdg ±3 digits

Full Scale Test Voltage: 20mV

11

4100 Series Digital u-Ohmmeter Specifications

11

4100ATC | 4150ATC | 4165 | 4165-1344