Atec Yokogawa-DR232 User Manual

Specifications

Advertising
background image

13

Standard Specifications

General Specifications

External Dimensions; Weight (with I/O module installed)

DR231: approximately 438 (W)

ϫ

291 (H)

ϫ

336 (D) mm; approximately 13 kg

DR232: approximately 438 (W)

ϫ

291 (H)

ϫ

301 (D) mm; approximately 9 kg

DS400: approximately 336 (W)

ϫ

165 (H)

ϫ

100 (D) mm; approximately 2.5 kg

DS600: approximately 422 (W)

ϫ

176 (H)

ϫ

100 (D) mm; approximately 3.5 kg

For model DR231, the DC power supply option adds 45 mm to the depth and 1.5
kg(f) to the weight.

AC Power Supply

Rated supply voltage:

100 to 240 VAC

Usable supply voltage:

90 to 250 VAC

Rated supply frequency:

50/60 Hz

DC Power Supply (/P6 option, only for the DR231 stand-alone model)

Rated supply voltage:

12 to 28 VDC

Usable supply voltage:

10 to 32 VDC

Terminal:

Dedicated connector

Note:

When both AC and DC power are connected to a DC
power supply model, which of the power supplies is
used depends on the voltage of the DC power supply
connected as follows.

DC Power Supply Voltage

<20 V

20 to 28 V

28 to 32 V

Power Supply Used

AC power supply

Indeterminate

DC power supply

Insulation Resistance

At least 20 M

at 500 VDC between the power supply and ground, between each

terminal and the ground, and between input terminals

Withstanding Voltage

Between power supply terminal and ground:

1,500 VAC (50/60 Hz, 1 min.)

Between input/output terminal and ground:

1,500 VAC (50/60 Hz, 1 min.)

Normal Operating Conditions

Supply frequency:

50 Hz

±

2% or 60 Hz

±

2%

Ambient temperature:

DR231, DR232 0 to 50˚C (FD operation 5 to 40˚C)

DS400, DS600 Panel mount –10 to 60˚C

Desk-top

–10 to 50˚C

Ambient humidity:

20 to 80% RH (between -10 and 40˚C)

Safety Standards

CSA C22.2 No. 1010.1-92, IEC1010-1:1995, EN61010

EMI Standard

EN55011:1991 Group 1 class A

EMC Standard

EN50082-2:1995

System Configuration

Configuration Method

DR231:

Configure a system with this model by specifying
necessary options, such as the input and
communications functions, according to the model
code when ordering.

DR232:

Configure a system with this model by combining one or
more of the modules and subunits listed below.

Connecting Modules and Subunits (DR232)

Standard Modules and Software for System Configuration

The following modules and software can be installed in a main unit and subunit to
configure a data acquisition system.
Input Modules:

Universal (DCV, TC, RTD and DI), DCV/TC/DI
dedicated, power monitor, strain, pulse, direct current
(mA) and digital input
Connectable to DS400 and DS600

Communications Modules:

Ethernet, GP-IB, RS-232C and RS-422A/485
Connectable to DR232 main unit

Alarm Contact Output Modules:

4 contacts (C contact: NO-C-NC) and 10 contacts (A
contact: NO-C)
Connectable to DR232 Main unit or DS400 and DS600

DI/DO Modules:

Two alarm output contacts (NO-C-NC) and fail output
Connectable to DR232 Main unit or DS400 and DS600
Up to 1 module/1 system can be connected.

Extension Modules:

Interfaces for remote power supply
One extension module can be connected to each
DS400 and DS600.
(should be used with extension base units)

Software:

DAQ 32(standard software)
DAQ 32 Plus (optional software)

Types and Number of Modules That Can Be Connected

DR231:

Specify the types of modules and the number
according to the model code.

DR232:

Communications module, DI/DO module or alarm
contact output module

DS400/600

Input module, alarm contact output modules, DI/DO
module and extension modules
Four or six modules can be connected.

Connection of Subunits

DR231:

Cannot be connected.

DR232:

Up to 6 subunits can be connected. One subunit can
be installed on the rear panel by screws.

Input Section

Number of Input Channels

DR231:

10 to 30 channels (Specify the number of channels
when ordering.)
Power monitor input option: 2 or 6 channels

DR232:

0 channel. Expandable up to 300 channels by
connecting subunits.

Types of Input Modules

DR231:

Universal (DC voltage, thermocouple, RTD and
contact), DCV/TC/DI dedicated (specify the types
when ordering), power monitor option

DR232:

Universal (DC voltage, thermocouple, RTD and
contact), DCV/TC/DI dedicated, power monitor,
strain, pulse, direct current (mA) and digital input
module

Measurement Range:

See the specifications for each input module.

Specifications

Measurement Interval:

0.5, 1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 and 60 seconds

DR231:

Maximum of 2 s per 30 channels

DR232:

Maximum of 500 ms per 300 channels (including the subunit)

The measurement interval is dependent on the slowest input module if input
modules of different measurement intervals are connected at the same time.

A/D Integration Period

Manual selection or automatic switchover between 20 ms (50 Hz), 16.7 ms (60
Hz) and 100 ms (10 Hz)
Minimum measurement interval when the 100-ms integration mode becomes:
DR231:

30 channels: 6 seconds

DR232:

4 seconds per 300 channels (including the subunit)
(depends on the modules and number of channels)

Recording section (DR231/232 main unit)

Recording Method

Raster scan method, 10-color wire dot recording

Number of Recording Points

300 points maximum (stand-alone model: 30 points + AC 6 points)

Recording Paper

Effective recording width:

250 mm (for analog trend measurement)

Analog recording color (You can specify a color for each channel.)

Purple, red, green, blue, brown, black, navy blue, yellow-green, red-purple,
orange

Analog Recording Interval

FIX:

Recording takes place at the specified measurement
interval between 2 and 60 seconds (not all measured
values are sampled for analog recording in case of
the 0.5- and 1-second measurement intervals)

AUTO:

Linked to recording paper feed speed

Recording Paper Feed

Paper feed speed:

1 to 1,500 mm/hour

Display Section

Display Section

Display:

VFD display (5 x 7 dot matrix, 3 lines)

Number of characters:

22 characters (large/1 line), 40 characters (2 lines)

Memory Function Section

Memory Media

3.5-inch floppy disk drive with 512 kB SRAM buffer memory

Data Capacity

10 data/ch to 50 kdata/ch
(Total data memory should be less than total memory length.)

Applicable data

Setting values, measured values and computed values except report calculating values

Memory Mode

Binary
Can be converted to ASCII (CSV) format for copying buffer memory data to floppy disk.

Sample Rate

Synchronized with the measurement interval of the recorder unit, or synchronized
with event.

Alarms

Number of Settings

Up to four settings can be made for each channel.

Kinds of Alarms

Upper/lower limit, difference upper/lower limit, upper/lower limit of percentage
change, upper or lower limit only for the results of computation
Percentage change alarm time interval: 1 to 15 scans

Number of Alarm Output Points

DR231:

12 maximum (alarm option: 10; DI/DO option: 2)

DR232:

300 in total

Standard Computation Functions

Kinds of Computation

Difference between arbitrarily selected channels, linear scaling, moving average, pulse integration
Scalable range:

DC voltage, thermocouple, RTD, contact

Scaling range:

–30,000 to +30,000

Moving average:

2 to 64 scans

Pulse integration:

Effective when a pulse input module is recognized
(up to 60 channels)

Fail, Chart End Output
(DR expandable model. The DR stand-alone model uses the /R1 option.)
Functions:

Refer to the DI / DO modules.

Optional Specifications

Computation Function (/M1)

Number of Computation Channels

DR231:

30 channels maximum

DR232:

60 channels maximum

Kinds

Remote RJC, four arithmetic operations, SQR (square root), ABS (absolute
value), LOG (common or natural logarithm), EXP (exponential), statistics
processing (CLOG, TLOG), logic (AND, OR, NOT, XOR), relative computation,
previous data reference
CLOG:

Mathematical processing within a group of data that
was measured at the same time (total, maximum,
minimum, average, max. - min.)

TLOG:

Mathematical processing of data from a certain
channel over a period of time (24 hours maximum)
(total, maximum, minimum, average max. - min.)

Report Function (/M3)
Instantaneous values of measured data, as well as maximum, minimum, average
and total, for each hour, day or month are printed in tabular form on recording
paper. Analog recording is interrupted while a report is being made.
Report calculation channels:

Up to 60 channels

Note:

This function does not allow the results of the report
and computing function to be saved on floppy disks.
(Thus, to be able to transfer the results to a personal
computer, the DP380 report software is needed.
Note that the DP380 software cannot be run
simultaneously with the DAQ32 or DAQ32Plus
software package.)

Power Monitor Options (/N7, /N8)

Applicable models and outline specifications

DR231 stand-alone model (For the DR232, the power monitor module is sold

Advertising