Atec Yokogawa_FX100 Series User Manual

Paperless recorder, Fx100 specification

Advertising
background image

NO

.

STYLE

MODEL

SUFFIX

SUPPL

Y

FREQ

U

ENCY

10Base-T

RS-422A/485

FG

SG

S

D

B

S

DA

R

D

B

R

DA

144.3 (5.68)

151.5 (5.96)

(0.95)

24.1

(2.30)

58.4

175 (6.89)

9.4

7.5

(0.37)
(min. space for mounting)

(0.30)
(dimension after mounting)

136.5

+0.4

0

(5.37)

144.3 (5.68)

151.5 (5.96)

(approx. inch)

Unit : mm

2 to 26 (panel thickness)

Ethernet

Power Supply
Terminal

Option Terminal

Input Terminal
1 to 12ch

RS-232 or RS-422A/485

Bulletin 04L20A01-00E

FX103/FX106/FX112

Paperless recorder

NOTICE

● The mounting hardware is attached on both sides or at the top and

bottom (2places) for panelmounting.

● See the general speelfication sheet (GS 04N03A01-00E) for the panel

cutout for
side by side or top-and bottom mounting.

www.yokogawa.com/tm/

... and subscribe to “Newswave,”
our free e-mail newsletter

*1: /A1, /A2, /A3 cannot be specified together.
*2: /C2, /C3 cannot be specified together.
*3: If /F1 is specified, /A3 cannot be specified.
*4: If /PM1 is specified, each /A3, /R1, and /M1 cannot be specified.
*4: If /PM1 is specified, /A2 and /F1 cannot be specified together.
In case that Modbus master function is utilized, either /M1 or /PM1 must be specified.
The TCP/IP software used in this product and the document for that
TCP/IP software are based in part on BSD networking software, Release 1 licensed from The
Regents of the University of California.
• Ethernet is a registered trademark of XEROX Corporation.
• Modbus is a registered trademark of AEG Schneider
• Other company and/or product names are registered trade mark of their manufactures.

Subject to change without notice.

[Ed : 01/b] Copyright ©2003

Printed in Japan, 310(YG)

YOKOGAWA ELECTRIC CORPORATION
Network Solutions Sales Business Div./Phone: (81)-55-243-0309, Fax: (81)-55-243-0397
E-mail: [email protected]

YOKOGAWA CORPORATION OF AMERICA

Phone: (1)-770-253-7000, Fax: (1)-770-251-2088

YOKOGAWA EUROPE B.V.

Phone: (31)-33-4641806, Fax: (31)-33-4641807

YOKOGAWA ENGINEERING ASIA PTE. LTD.

Phone: (65)-62419933, Fax: (65)-62412606

RS-12E

FX100 Specification

Input section

Measurement points:

FX103 3 (three) channels
FX106 6 (six) channels
FX112 12 (twelve) channels

Measurement interval: FX103 250 ms

FX106, FX112 1s, 2s

Input Types

DCV:20mV, 60mV, 200mV, 2V, 6V, 20V, 50V
TC: R, S, B, K, E, J, T, N, W, L, U, WRe
RTD: Pt100, JPt100 (Pt1000: Optional)
DI:

voltage input (TTL), contact input

DCA:with external shunt register

Simple Calculation:

Differential, linear scaling, square root

Maximum allowable input voltage:

±

10 VDC (continuous) for less than 200mVDC range, TC, RTD, DI

ranges

±

60 VDC (continuous) for more than 2 VDC range

Maximum common mode noise voltage:

250 Vrms AC (50/60Hz)

Maximum noise voltage between channels:

250 Vrms AC (50/60Hz)

Display

Display:

5,5“ TFT color-LCD

Display types:

Trend, bar graph, digital, overview, historical,
information (alarm, message, memory)

Language:

English, French, Germany, Japanese

Data Storage Function

Internal memory:

Flash memory

External media:

One of the following external media options can be specified:
- None, - 3,5“ floppy disk, - compact flash memory card (32MB-512MB)

Saving method:

Automatic saving or manual saving

Data type:

Display data, event data

Data format:

Binary format

Alarm Function

Number of levels:

4 levels per channels

Type of alarm:

Higher/lower limits, differential higher/
lower limits, rate-of-change rising/falling limits

Construction

Dimensions:

144(W) x 144(H) x 234(D) mm

Weights:

FX1xx-0: approximately 2.2 kg
FX1xx-1: approximately 2.5 kg
FX1xx-4: approximately 2.3 kg

Front panel:

Water and dust proof (based on IEC529-IP65, except external icing
test, except for side-by-side mounting)

Power Supply

Rated voltage:

100 to 240 VAC

Rated frequency:

50/60 Hz

Power consumption:

normal 22 VA, maximum 35 VA

Safety and EMC Standards

Safety standards:

Certified by CSA22.2 No.1010-1, UL61010B-B complies with
EN61010-1

EMC standards:

Complies with EN61326-1

Optional Pulse Input Functions (/PM1)

Number of channels:

3 for pulse, and 5 for remote controlling

Applicable frequency:

Up to 100 Hz

Minimum pulse width detection:

Low (close), high (open) : more than 5ms

Optional Relay Output (/A1, /A2, /A3)

Number of outputs:

2,4,or 6 points; energized/
di-energized, hold/non-hold, AND/OR

Optional Serial Communication functions (/C2, /C3)

Connection:

EIA RS232 (/C2) or RS-422-A/485 (/C3)

Protocols:

Yokogawa private protocol

Modbus function:

MODBUS RTU MASTER/ SLAVE

Optional Ethernet Communication functions (/C7)

Connection:

Ethernet (10Base-T)

Functions:

E-mail notification, Web-server, FTP function

Optional Mathematic Functions (/M1)

Number of channels:

Up to 12 channels

Operations:
General arithmetic operations:

Four arithmetic operations, square root, absolute, common
logarithm, exponential, power, relations, logic operations.

Statistical operations: Average, minimum, maximum, and summation
Constant:

Available (up to 30 constants)

Digital data input via communication:

Up to 12 data

Report function:
Type:

Hourly, daily, hourly+daily, daily+weekly, daily+monthly

Operation:

Average, maximum, minimum, summation

Data format:

ASCII

Optional Remote Controlling Functions (/R1)

Number of channels:

8 points

Operable items:

Memory start/stop, computation start/stop, computation data reset,
alarm acknowledgement, event file external trigger, message writing,
load setting, time setting, snapshot.

FX103

FX106

FX112

External memory -0

-1

-4

Display Language

-2

Option

/A1

/A2

/A3

/C2

/C3

/C7

/F1

/M1

/N2

/N3

/PM1

/R1

FX100(3ch)

FX100(6ch)

FX100(12ch)

None

FDD (3.5 inch, 1.44MB)

Compact flash memory card(32MB)

English (German/French)

Alarm output 2 points *

1

Alarm output 4 points *

1

Alarm output 6 points *

1

RS-232 communication interface *

2

RS-422-A/485 communication interface *

2

Ethernet(10BASE-T) communication interface

FAIL/memory end output *

3

Mathematical function (with report function)

3 legs isolated RTD

Pt1000

input

Pulse input 3 points, remote control 5 points (including Mathematical function )*

4

Remote control 8 points

Model

Description

Suffix

code

Option

code

Advertising