KeyTek ECAT Models EP61 and EP62

Plug-in modules provide swells, dips and interrupts on AC
power mains in compliance with, and exceeding the
requirements of IEC 61000-4-11 Edition 2. Model EP61 for
single-phase AC lines to 240 RMS, 16A; Model EP62 for
single-phase AC lines to 240 RMS, 32A

AC INPUTS/OUTPUTS

Input Voltage for 100%

50 to 240V at 50Hz and to 277V at 60Hz

Output Voltages on

0% (open or short), 40%, 50%, 70%,

the Selected Phase

80%, 90%, 100%, 110%, 120% and 150%

EP61 Output Current

16A at 250V; 20A at 125V*

EP62 Output Current

32A at 250V; 30A at 125V*

*The actual AC mains voltage and current limit is based on the mains connector selected.

Inrush Current

>250A at 120V; >500A at 220-240V

Event Duration

From 0.03 cycle (10°) to 500 minutes;
maximum 12 events per cycle

Switching Times

1-5µs into a 100 ohm load

Overshoot

<5%

Undershoot

<5%

MEASUREMENTS

rms Voltage

0-300V, 0.5% of range + 1% of reading

rms Current

0-40A, 0.5% of range + 1% of reading

Peak Current

0-1000A, 1% of range + 5% of reading

Inrush Current Qualification

Internal, built-in circuit according to
IEC 61000-4-11. Automatically measures peak
inrush current at 90° and 270°. Peak values are
reported via the control software.

Minimum System Requirements: E100 series control center

KeyTek ECAT

®

Model E455x

Single and three-phase AC line coupler/decouplers for EFT
and Surge waves, as specified by IEC 61000-4-4 Edition 2
and IEC 61000-4-5

ELECTRICAL
Model

Single or Three-phase

Voltage

Current** per phase

E4551A/E4551kV*

Single-phase

250V rms

15/16A***

E4552A/E4552kV*

Single-phase

277V rms

32A

E4553A/E4553kV*

Three-phase

480V rms

16A

E4554A/E4554kV*

Three-phase

480V rms

32A

E4555

Three-phase

600V rms

50A

E4556

Three-phase

600V rms

100A

* kV version is required for operation with surge modules greater than 7kV, such as the
E510A. All standard coupler/decoupler options apply

** Actual current capability may be limited by the AC line connectors selected

*** Depends on connector selected. Typically 15A with U.S. NEMA connector; 16A with
appropriate European style connectors

Coupling Mode

Coupling mode selection is controlled manually from the control
selection center, or automatically using KeyTek SurgeWare

or

BurstWare

software. Coupling is allowed from any line to any

other line or combination of lines.

Monitoring

Monitoring and peak detection of surge voltage across any two
manually-selected lines. Monitoring can be at the EUT or at the
front panel of the coupler/decoupler.

COMBINED SURGE & EFT COUPLER/DECOUPLERS

Monitoring and peak detection of surge current in either High or Neutral, selected by the
ECAT Control Center or the computer, measured without including back-filter surge current.

Minimum

E100 series control center EFT or mains - coupled surge module

Requirements

Options

E455x-DC

Allows the E455x coupler/decoupler to be used with DC as well as

AC

mains. The DC current ratings for essentially resistive loads are:

to 48V

to 110 V

to 220V

E4551A/E4551kV

15A

5A

0.8A

E4552A/E4552kV

15A

5A

0.8A

E4553A/E4553kV

20A

8A

1.2A

E4554A/E4554kV

25A

8A

1.2A

E4555

50A

50A (120V)

30A

E4556

100A

50A (120V)

30A

E455x-VI

Enhanced V and I monitoring. Adds monitoring and peak detection of
surge voltage and current. Upper and lower limits can be placed on
surge peaks. Monitoring of 3 wires is provided in single-phase
systems, 5 wires in three-phase systems. Selection of the V and I
inputs is performed from the control center or can be made
automatically with SurgeWare control software

E455x-HV

Increases the AC mains voltage rating from 277V to 480V rms in the
E4552, and from 480V to 600V rms in the E4553 and E4554. The HV
option is not available in the E4551, E4555 and E4556.

Physical

Physical size of module varies depending on model number

PQF

(POWER QUALITY FAILURE) MODULES

Advertising