background image

All test parameters under full software control via

KeyTek BurstWare

:

Polarity

Burst Voltage

0-360° Line Sync

Frequency

Burst Duration

Coupling Modes

Period Test

Duration

Burst Voltages from 200V to 4.4kV

IEC specified 5ns/50ns impulses into 50 ohms- also

meets the proposed 5/100ns requirement into a high
impedance.

50 ohm source impedance

Built-in pulse monitor, attenuated to 1kV/V

for convenient monitoring without the need for
high voltage probes.

Real World EFT options including Chirp and

frequencies to 2MHz.

ECAT

®

Model E411

EFT/Noise

Burst Simulator for

IEC 1000-4-4 to 4.4 kV

A plug-in module for ECAT
systems for testing in accordance
with IEC 1000-4-4 (IEC 801-4).

FEATURES

The ECAT is configurable as a compact,

standalone system with one test capability or as
an integrated, multiple capability tester.

Unparalleled safety features for high voltage

testing, including a complete interlock system that
not only disables simulator operation if activated, but
completely removes AC mains power from the EUT,
when using any KeyTek AC mains coupler/decoupler.

Automatic report generation using KeyTek’s

exclusive Windows

®

-based software.

Modular system construction provides the ability to

upgrade test capabilities as requirements change or
evolve.

Single port testing for Surge, EFT and PQF

mains

coupled disturbances allows changing test modes
without switching off power to the EUT. This avoids
lost time due to re-initializing the EUT between tests.