Atec Yokogawa-700929 User Manual

Atec Equipment

Advertising
background image

YCA T&M HFI SM 11/99

Yokogawa Corp. of America

Probe Data Sheet
Model

700929

30 MHz 10:1 Passive Probe for Isolated BNC inputs

-------------------------------------------------------------------------------------------------------------

Overview

The 700929 is an isolated passive probe with an attenuation ratio of 10:1. It designed to be used with isolated input
connectors.
The 700929 is the only probe to be used with the 700926 isolator. The 700929 is recommended for use with the
701855 High Speed Isolation Module of the DL700 series.


Specifications

Item

Specification

Bandwidth

DC

to

250

MHz

Attenuation

ratio

10:1

+/-

1.5%

Input impedance/capacitance

10 M

+/- 1%, approx. 15 pF

Rise time

1 ns or less

Maximum input voltage (probe only)

1000 Vrms, 1 MHz or lower

Maximum input voltage (with 700926)

600 Vrms, 160 kHz or less

Connector

type

Isolated

BNC

Total

length

1.5

m


Probe Items

Contents of the package consists of the probe 700929, a pincher tip and an adjustment screwdriver.

Probe

Pincher

tip

Adjustment

700929 B99440WH

screwdriver

Advertising