Atec Tektronix-TDS3000 Series User Manual

Page 2

Advertising
background image

T D S 3 0 0 0 S E R I E S D I G I T A L P H O S P H P O R O S C I L L O S C O P E S

2

QUICK TO LEARN AND QUICK TO USE
The TDS 3000 Series graphical user inter-
face offers a new operating mode called
QuickMenu. This quick access user inter-
face makes the main oscilloscope controls
accessible with a push of a single button.
Included with every scope is an On-Line
Tour disk that runs in the oscilloscope.
This disk gives an overview of the prod-
uct’s operation and capabilities.

FLEXIBLE FEATURES FOR EVERY
APPLICATION
The portable form factor allows the
instrument to go wherever it is needed.
The Communication and Application
Modules enable the instrument to be con-
figured for specific applications or
upgraded without returning it to the man-
ufacturer. The TekProbe

®

Level II

interface provides power to a range of
application specific accessories.

APPLICATION MODULES
With the application module concept, the
TDS 3000 Series can be easily adapted to
the needs of the user. These modules are
easily installed by the user. At power on
the scope indicates which modules are
installed. Currently, there are five applica-
tion modules available for the TDS 3000
Series – a telecommunications mask test
module, an FFT module, an advanced
trigger module, a 601 digital video module
and an extended video module. In addi-
tion, there are three communication
modules available – a 10Base-T LAN/RS-
232 module, a GPIB/RS-232 module and a
VGA/RS-232 module. Centronics port is
standard.

ACQUISITION MODES

Peak Detect – High frequency and random
glitch capture. Captures glitches as narrow as
1 ns.
Sample – Sample data only.
Envelope – Max/Min values acquired over one
or more acquisitions.
Average – Waveform data from 2 to 572
(selectable) acquisitions is averaged.
Single Sequence – Use SINGLE SEQUENCE
button to capture a single triggered acquisition
sequence at a time.

TRIGGER SYSTEM

Main Trigger Modes – Auto (supports Roll
Mode for 40 ms/div and slower), Normal.
B Trigger – Trigger after time or events.
Trigger After Time Range – 13.2 ns to 50 s.
Trigger After Events Range – 1 to 9,999,999
events.
External Trigger Input (available on TDS 30X2
only) >1 M

in parallel with 17 pF; Max input

voltage is 150 V RMS.

TRIGGER TYPES

Edge – Conventional level-driven trigger. Positive
or negative slope on any channel. Coupling selec-
tions: DC, noise reject, HF reject, LF reject.
Video – Trigger on all lines or individual line,
odd/even or all fields, or analog HDTV formats
(1080i, 1080p, 720p, 480p). See optional
TDS 3VID and TDS 3SDI application modules for
extended video triggering and measurement
features.
Logic (Standard on TDS 30X4, must purchase
TDS 3TRG for TDS 30X2)
PATTERN: Specifies AND, OR, NAND, NOR when
true or false for a specific time.
STATE: Any logic state. Triggerable on rising or
falling edge, of a clock.
Note: Logic triggers can only be used on combi-
nations of 2 inputs.
Pulse (Standard on TDS 30X4, must purchase
TDS 3TRG for TDS 30X2)
WIDTH (or GLITCH): Trigger on pulse width less
than, greater than, equal to, or not equal to a
selectable time limit ranging from 39.6 ns to 50s.

RUNT: Trigger on a pulse that crosses one thresh-
old but fails to cross a second threshold before
crossing the first again.
SLEW RATE: Trigger on pulse edge rates that are
either faster or slower than a set rate. Edges can
be rising, falling, or either.
Comm (must purchase TDS 3TMT) provides
isolated pulse triggering required to perform
DS1/DS3 telecommunications mask testing per
ANSI T1.102 standard.

MEASUREMENT SYSTEM

Automatic Waveform Measurements –
Period, Frequency, +Width, –Width, Rise Time,
Fall Time, +Duty Cycle, –Duty Cycle,
+Overshoot, –Overshoot, High, Low, Max, Min,
Pk-Pk, Amplitude, Mean, Cycle Mean, RMS, Cycle
RMS, Burst Width.
Display any four measurements from any combi-
nation of waveforms.
Thresholds – Settable in percentage or voltage.
Gating – Measurements can be gated using the
screen or vertical cursors.

Characteristics

T D S 3 0 0 0 S e r i e s E l e c t r i c a l C h a r a c t e r i s t i c s

TDS 3012

TDS 3032

TDS 3052

TDS 3014

TDS 3034

TDS 3054

Bandwidth

100 MHz

300 MHz

500 MHz

100 MHz

300 MHz

500 MHz

Channels

2

2

2

4

4

4

Sample Rate on Each Channel

1.25 GS/s

2.5 GS/s

5 GS/s

1.25 GS/s

2.5 GS/s

5 GS/s

Maximum Record Length

10K points on all models

Vertical Resolution

9-bits on all models

Vertical Sensitivity (/div)

1 mV-10 V on all models

Vertical Accuracy

±2% on all models*

Max Input Voltage (1 M

)

150V RMS CAT I on all models

Position Range

± 5 div on all models

BW Limit

20 MHz

20, 150 MHz

20, 150 MHz

20 MHz

20, 150 MHz

20, 150 MHz

Input Coupling

AC, DC, GND on all models

Input Impedance Selections

1 M

in parallel with 13 pF, or 50

Time Base:Range (/div)

4 ns – 10 s/div

2 ns – 10 s/div

1 ns – 10 s/div

4 ns – 10 s/div

2 ns – 10 s/div

1 ns – 10 s/div

Accuracy

200 ppm

200 ppm

200 ppm

200 ppm

200 ppm

200 ppm

Display Monitor

Color LCD

* Derated at 0.07%/°C for temperatures above +28°C and below +18°C.