For 100A : 1A and 1000A : 1A AC Current Measurements

CT1000 Current T

ransformer

Specification

Frequency Range:

45Hz to 1kHz

Current Range:

100:1 ratio

10A to 120A rms

1000:1 ratio

100A to 1200A rms

Maximum Input Current:

1000A continuous
2000A for 1 hour

Accuracy (23°C ± 5°C):

±0.2% of specified ratio

Phase Error (23°C ± 5°C):

Better than ±0.1° at 50Hz

Maximum Burden:

0.15VA for specified accuracy (equivalent to the input of

any Voltech power analyzer plus the test leads provided)

Dielectric Strength:

4kV at 50Hz for 1 minute

Temperature Range:

Operating:

-10°C to +50°C

Storage:

-25°C to +85°C

The Voltech CT1000 is a toroidally wound AC current transformer that extends the range of Voltech

power analyzers such as the PM100, PM300 or PM3000A to 1200A rms. Close- tolerance current

ratios provide a measurement accuracy of better than 0.2%, making the CT1000 suitable for preci-

sion-grade measurements.

The CT1000 is housed in a fully insulated, shock-resistant casing with a core aperture that accom-

modates large-diameter cables or rectangular busbars. To minimize the risk of electrical shock, the

CT1000 has 4mm safety sockets for the leads connecting it to the power analyzer. In addition, a

novel over-voltage protection circuit limits the voltage on the output terminals to 50V peak, if the

transformer’s secondary winding is inadvertently open-circuited.