Highest Performance-to-Cost ratio in the industry
Up to 2000V peak and 150A peak measurable internally
(external CT capable)
Two channels for independent or synchronous measurements,
including input/output effi ciency and loss (2802 only)
Base accuracy <0.08%. Current and voltage accuracies specifi ed
to less than 1mArms and 1Vrms respectively (<0.2%)
Measures and displays volts, current, power, frequency, harmonics
(to the 100th), THD, PF, CF, K-Factor, Triplens, inrush, distortion,
glitches, and much more
Provides PASS/FAIL tests to user limits*
Integrated line switch and inrush waveforms
Source or load measurements with wiring loss and voltage
burden compensation
DC charge and discharge
measurements
Frequency Range: DC and
20mHz – 200kHz
Graphics display shows
numerical results, wave-
forms, bar graphs, startup &
history charts with zoom & scroll features
16-bit A/D converters at 235ksps
12 User-confi gurable digital I/O
USB Flash drive support for data logging*
Communications interfaces include GPIB (IEEE488), RS-232, USB
(host and device), and Ethernet*
Suitable for AC, DC, 1-phase 3-wire, 2-phase 3-wire

, 3-phase 3-wire

,

in/out synchronous

, or independent measurements (

2802 only).

»
»

»

»

»

»
»
»

»

»

»

»
»
»
»

»

9770-A Carrol Centre Road | San Diego, CA 92126 | 858.530.8099 phone | 858.530.8077 fax

www.xitrontech.com | [email protected] | [email protected]

2801 and 2802 Advanced Power Analyzers

The 2801 and 2802 Advanced Single- and Two-Channel
Power Analyzers combine an unprecedented feature
set with an ideal combination of precision, speed
and ease-of-use in an instrument so economical
it can be on every bench.

With an extended measurement range from micro-amps to hundreds
of amps and millivolts to kilovolts, the 2801 and 2802 are the ideal
analyzers for standby power or Energy Star testing. In addition to
numerical results, the 2801 and 2802 captures waveforms with true
512-point precision. Results and waveforms can be displayed, read
via the communication ports, or sent directly at full resolution to a USB
printer. Power and amplitude measurements with an accuracy of <0.1%
are automatically synchronized to the fundamental frequency. Peak,
rms, rectifi ed, and DC measurements of voltage, current and power are
provided including continuous, inrush, startup and history modes, plus
an integration mode for W-Hr, A-Hr, VA-Hr, as well as integrated average
power. (Channels can be operated synchronously or independently on
the 2802) The 2801 and 2802 also provide waveform peak and glitch
capture modes.

Quality and Reliability

XiTRON Technologies, founded in 1990, is the premier source of
precision power testing and measuring equipment for industrial and
consumer product development and manufacturing. XiTRON’s
sophisticated technology provides companies the edge in design
verifi cation and product manufacturability. XiTRON is ISO9001-2000
certifi ed, EN46001 registered and FDA (GMP820) compliant.

XiTRON XVIEW Software

While all XiTRON precision test equipment is designed to

be used in a completely stand-alone manner, there are

times when external tools can aid or enhance the operation

of an instrument. XView software tools and drivers are designed to help
easily confi gure an instrument from a single screen, or are used to view a
complete set of measurements in a single screen. Other XView tools are
designed for data collection where results can be recorded in an Excel-
compatible fi le for post-processing, insertion into reports, or simply for
archival purposes.

These precompiled applications can be downloaded from the XiTRON
web site at

www.xitrontech.com/support.html.

Often, XiTRON can provide

the source code for a particular application, and it can be used as a
convenient ‘starting point’ for a custom software application.

(*planned)