B&K Precision 2712 - Manual User Manual

B&K Precision Tools

Advertising
background image

OPERATING INSTRUCTIONS

MODEL 2712

DIGITAL MULTIMETER

SAFETY INFORMATION

The following safety information must be observed to ensure maximum personal

safety during the operation at this meter:

Use the meter only as specified in this manual or the protection provided by the

meter might be impaired.

Test the meter on a known voltage before using it to determine if hazardous

voltage is present.

Do not use the meter if the meter or test leads look damaged, or if you suspect

that the meter is not operating properly.

Never ground yourself when taking electrical measurements. Do not touch

exposed metal pipes, outlets, fixtures, etc., which might be at ground potential.

Keep your body isolated from ground by using dry clothing, rubber shoes, rubber

mats, or any approved insulating material.

Turn off power to the circuit under test before cutting, unsoldering, or breaking

the circuit. Small amounts of current can be dangerous.

Use caution when working above 60V dc or 30V ac rms. Such voltages pose a

shock hazard.

When using the probes, keep your fingers behind the finger guards on the

probes.

Measuring voltage which exceeds the limits of the multimeter may damage the

meter and expose the operator to a shock hazard. Always recognize the meter

voltage limits as stated on the front of the meter.

SPECIFICATIONS

Display: 4¾ digit liquid crystal display (LCD) with a maximum reading of 40000.

Analog bargraph: 40 segments with measurements 20 times per second.

Polarity: Automatic, positive implied, negative polarity indication.

Overrange: MSD(Most Significant Digit) blinks .

Zero: Automatic.

Low battery indication: The " " is

displayed when the battery voltage drops

below the operating level.

Measurement rate: 2 times per second, nominal.

Auto power off: Approx. 30 minutes.
Operating environment: 0℃ to 50℃ at < 70% relative humidity.
Storage temperature: -20℃ to 60℃ at < 80% relative humidity.
Accuracy: Stated accuracy at 23℃±5℃, <

75% relative humidity.
Temperature Coefficient: 0.1 x (specified accuracy) per ℃. (℃ to 18℃, 28℃
to 50

℃).

Altitude: 6561.7 feet (2000m).

Power: Single standard 9-volt battery, NEDA

1604, JIS 006P, IEC 6F22.

Battery life: 150 hours typical with carbon-zinc.

Dimensions: 165mm (H) x78mm (W) x42.5mm (D).

Weight: Approx. 10.0 oz.(285g) including holster.

Accessories: One pair test leads, one spare

fuse, 9V battery (installed), and Operating

Instructions.

DC VOLTS

Ranges: 400mV, 4V, 40V, 400V, 1000V

Resolution: 0.01mV

Accuracy: ± ( 0.15% rdg + 10 dgts) on 400mV range

± ( 0.1% rdg + 5 dgts) on 4V to 1000V ranges

Input impedance: 400mV: >3MΩ; 4V ~

1000V:2.3MΩ

Overload protection: 1000VDC or 750VAC rms

AC VOLTS

(TRUE RMS) (45Hz – 1kHz)

Ranges: 400mV, 4V, 40V, 400V, 750V

Resolution: 0.01mV

Accuracy:

± ( 1.2% rdg + 20dgts) 45 ~ 60Hz on 400mV to 400V ranges

± ( 1.5% rdg + 20 dgts)

60 ~ 500Hz on 4V range

± ( 1.5% rdg + 20dgts) 60 ~ 1kHz on 40V to 400V ranges

± ( 2.0% rdg + 20 dgts)

45 ~ 500Hz on 750V range

AC+DC VOLTS

(TRUE RMS) (45Hz - 1kHz)

Ranges: 400mV, 4V, 40V, 400V, 750V

Resolution: 0.01mV

Accuracy:

± ( 1.5% rdg + 20dgts) 45 ~ 60Hz on 4V to 400V ranges

± ( 2.0% rdg + 20 dgts)

60 ~ 500Hz on 4V range

± ( 2.0% rdg + 20dgts) 60 ~ 1kHz on 40V to 400V ranges

± ( 2.0% rdg + 20 dgts)

45 ~ 500Hz on 750V range

Crest factor: ≤3

AC coupled true rms specified from 10% to 100% of range.

Input impedance:

400mV: >3MΩ; 4V ~750V:2.3MΩ

Overload protection:

1000VDC or 750VAC rms

CURRENT

Ranges: 40mA, 400mA, 10A

Resolution: 1uA

DC accuracy:

± ( 0.5% rdg + 10 dgts) on 40mA to 400mA ranges

± ( 1.5% rdg + 10 dgts) on 10A range

AC accuracy: (TRUE RMS) (50Hz ~ 1kHz)

± ( 2.0% rdg + 30 dgts) on 40mA to 400mA ranges

± ( 2.5% rdg + 30 dgts) on 10A range

Crest factor: ≤3

AC coupled true rms specified from 10% to 100% of range.

Voltage burden: 0.2V on 40mA, 10A ranges

1V on 400mA range

Input protection: 0.5A/500V fast blow ceramic fuse

10A/500V fast blow ceramic fuse

10A Input: 10A for 60 seconds maximum followed by a 10 minute cooling period.

RESISTANCE

Ranges: 400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, 40MΩ

Resolution: 0.01Ω

Accuracy:

± ( 0.3% rdg + 15dgts) on 400Ω range

± ( 0.3% rdg + 5dgts) on 4kΩ to 400kΩ ranges

± ( 0.5% rdg +10 dgts) on 4MΩ range

± ( 1.5% rdg + 20 dgts) on 40MΩ range

Open circuit volts typical: 1.2Vdc (2.5Vdc on 400Ω range)

Overload protection: 500VDC or AC rms

CAPACITANCE

Ranges: 4nF, 40nF, 400nF, 4uF, 40uF

Resolution: 1pF

Accuracy: ± ( 3.0% rdg + 20 dgts) on 4nF range

± ( 3.0% rdg + 5 dgts) on 40nF to 400nF ranges

± ( 3.0% rdg + 5 dgts) on 4uF to 20uF ranges

± ( 5.0% rdg + 5 dgts) on 20uF to 40uF ranges

Overload protection: 500VDC or AC rms

FREQUENCY

Ranges: 100Hz, 1kHz, 10kHz, 100kHz, 500kHz

Resolution: 0.01Hz

Accuracy: ± ( 0.1% rdg + 10 dgts)

Sensitivity: 5Hz ~ 100kHz: >500mV rms, 100kHz ~ 500kHz: >1.5V rms

Minimum input range: >5Hz

Minimum pulse width: > 2us

Duty cycle limits: > 30% and < 70%

Overload protection: 500VDC or AC rms

DIODE TEST

Test current: 0.8mA (approximate)

Accuracy: ±(1.5% rdg + 10 dgts)

Open circuit volts: 3.0Vdc typical

Overload protection: 500VDC or AC rms

CONTINUITY

Audible indication: Less than 40Ω

Response time: 500ms

Overload protection: 500VDC or AC rms

fuse on 1.5V range.

OPERATION

Before taking any measurements, read the Safety Information Section. Always

examine the instrument for damage, contamination (excessive dirt, grease, etc.)

and defects.

Examine the test leads for cracked or frayed insulation. If any abnormal

conditions exist do not attempt to make any measurements.

MAX / MIN/PEAK

The "MAX" displays the maximum value of measurements. The "MIN" displays

the minimum value of measurements. press MAX/MIN/PEAK button for more

than 2 seconds on this button allows to switch PEAK mode. Press

MAX/MIN/PEAK button again to exit the mode.

PEAK mode: It is usable with AC voltage, AC current measurements.

PEAK mode response time: more than 1 ms.

,

,

Button

Use this button to select

or ,

or

,

or

range

position.

HOLD

Press [HOLD] button to toggle in and out of the Data Hold mode.

In the data hold mode, the " HOLD" annun-ciator is displayed and the last

reading is held on the display.

Press [HOLD] button again to release the hold and current readings are once

again displayed.

Backlight

Press the button to activate the backlight for approximately 60 second.

Manually Selecting Range

The meter also has a manual range mode. In manual range, you select and lock

the meter in a range. To manually select a range:

Press [RANGE] button to hold the selected

range. Subsequently pressing the [RANGE] button will select each range in

sequence from the lowest to highest range. Hold the button for 2 seconds to

return to the Autorange Mode.

Voltage Measurements

1.Connect the red test lead to ”VΩ” jack and

the black test lead to the ”COM” jack.

2.Set the Function/Range switch to the desired voltage type (AC or DC) and

range. If magnitude of voltage is not known, set switch to the highest range

and reduce until a satisfactory reading is obtained.

3.Connect the test leads to the device or circuit being measured.

4. For dc, a (-) sign is displayed for negative polarity; positive polarity is implied.

Current Measurements

1.Connect the red test lead to the (mA or 10A) jack and the black test lead to the

"COM" jack.

2.Set the Function/Range switch to the DC or AC ranges.

3.Remove power from the circuit under test and open the normal circuit path

where the measurement is to be taken. Connect the meter in series with the

circuit.

4.Apply power and read the value from the display.

Resistance Measurements

1.Set the Function/Range switch to the desired resistance range.

2.Reove power from the equipment under test.

3.Connect the red test lead to the "VΩ" jack and the black test lead to the "COM"

jack.

4.Connect the test leads to the points of measurements and read the value from

the display.

Diode Tests

1.Connect the red test lead to the “VΩ” jack and the black test lead to

the ”COM” jack.

2.Set the Function/Range switch to the “

“ position.

3.Turn off power to the circuit under test. External voltage across the

components causes invalid readings.

4.Touch probes to the diode. A forward-voltage drop is about 0.6V (typical for a

silicon diode).

5.Reverse probes. If the diode is good, “MSD blinks”

displayed. If the diode is shorted, “000” or another number is displayed.

6. If the diode is open, “MSD blinks”.

Continuity Measurements

1.Set the Function switch to the

position.

2.Turn off power to the circuit under test. External Voltage across the

components causes invalid reading.

3.Connect the test leads to the two points at which continuity is to be tested. The

buzzer will sound if the resistance is less than approximately 40Ω.

Capacitance M easurements

1.Set the Function/Range switch to the desired capacitance range.

2.Connect the red test lead to the “VΩ” jack and the black test lead to the “COM”

jack.

3. Touch the probes to the capacitor. Observe polarity when measuring polarized

capacitors.

4.Read the capacitance directly from the display.

5. Discharge the capacitor before taking capacitance measurements.

Frequency M easurements

1.Set the Function/Range switch to the

”Hz” position.

2. Use this button to select “Hz “ range position.

3.Connect the red test lead to the “VΩ” jack and the black test lead to the

“COM” jack.

4.Connect the test leads to the point of measurement and read the frequency

from the display.

Auto Power Off

1. Auto power off: approx. 30 minutes.

2. After auto power off, move the function switch to OFF position to re-start the

meter.

MAINTENANCE

WARNING

Remove test leads before changing battery or fuse or

performing any servicing.

Battery Replacement

Power is supplied by a 9 volt battery. (NEDA 1604, IEC 6F22). The "

" appears

on the LCD display when replacement is needed. To replace the battery, remove

the three screws from the back of the meter and lift off the front case. Remove

the battery from case bottom.

Fuse Replacement

If no current measurements are possible. Check for a blown overload protection

fuse. For access to fuses, remove the three screws from the back of the meter

and lift off the front case. Replace F1 only with the original type 0.5A/500V, fast

acting ceramic fuse, 6.35x32mm Replace F2 only with the original type

10A/600V, fast acting ceramic fuse, 6.35x25.4mm.

Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or

solvents. Dirt or moisture in the terminals can affect readings.

Safety: Conforms to IEC61010-1 (EN61010-1), CATII 1000V, CATIII 600V, Class

II, Pollution degree 2 Indoor use.

CATII: Is for measurements performed on circuits directly connected to the

low-voltage installation.

CAT III: Is for measurements performed in the building installation.

EMC: Conforms to EN61326.

The symbols used on this instrument are:

Caution, refer to accompanying documents

Equipment protected throughout by Double insulation (Class II)

Alternating current

Direct current

Ground

Advertising