3 width sonar (option), 2 diameter gauge – CEMB USA ER80 (D) User Manual

Page 18

Advertising
background image

18

Calibro a riposo

Piano flangia

0 Sonar

A

Use and maintenance manual Rev. 11-2012

ENGLISH

6.5.2 Diameter gauge

E

Display the SETUP menu

1. Press

/

to view the diameter gauge

CALIBRATION function.

m

2. Place the gauge rod on the spindle shell as shown

in the figure and press

3. Turn the gauge downward positioning the gauge rod

in contact with the spindle sleeve as shown in the

figure

and press

CALIBRATION COMPLETE

▪ Return the gauge to rest position.

▪ The wheel balancer is ready for operation.

INdIcATIoN

In case of errors or malfunctions, the indication of the
same step [P.1] or [P2], always reappears on the display.
Move the gauge back into rest position and repeat the
calibration operation as described above; if the error

persists, contact Technical Service.

If erroneously accessing the diameter gauge calibration

function, press

to cancel it.

6.5.3 Width sonar (option)

Display the SETUP menu

1. Press

/

to view the width sonar

CALIBRATION function.

2. Set with

/

the distance in mm betwe

en the Sonar sensor (0 sonar) and t

he end of the

distance gauge (at rest).

A = Distance: Gauge at rest to

0 sonar

In the event of incorrect input in the width gauge calibration

function, press

to cancel it.