Com-Power CDN-AF4E User Manual

Features, Application, Description

Advertising
background image

Features

Frequency Range 150 MHz to 230 MHz

Designed for IEC / EN 61000-4-6

Four wire unscreened, unbalanced cables

Individual calibration included

Three Year Warranty

Application

During conducted Immunity testing, CDNs are

utilized to provide a means of coupling RF common

mode signals to each line. In addition, CDNs provide

the required common mode impedance to the EUT,

isolation to the auxiliary equipment via common

mode decoupling of the disturbance signals and

provide uninterrupted communication between

the EUT and auxiliary equipment.

Before you begin testing with the CDN-AF4E

you will need to establish calibrated drive levels

corresponding to your desired test levels. During

drive level calibration, the RF signal level being

injected to the CDN is adjusted incrementally until

the voltage level measured at the 150Ω to 50Ω

adapter (ADA-515-2) connected to the EUT port is

approximately equal to the Umr value given in the

table below. The ADA-515-2 and accessories needed

for this test are available from Com-Power.

Test Levels

Open Circuit Voltage

Umr

1

0.167

3

0.5

10

1.67

Umr= Voltage level measured at the output of the 150Ω to 50Ω

adapter (ADA-515-2)

Description

Com-Power CDN-AF4E is a part of the series of

Coupling/Decoupling Networks designed specifically

for testing product for conducted immunity per

IEC / EN 61000-4-6.

The CDN-AF4E is designed for testing products

that uses four wire unscreened cables for data

communication. It has four 2 mm shrouded banana

sockets for both EUT and AE power connection.

The CDN-AF4E can handle up to 5 Amps of current.

The RF disturbance signal coupling port is female

BNC. It can handle up to 40V of RF Input Voltage.

The bottom surface of the CDN is not painted for

easy and effective grounding.

All Com-Power CDNs are individually calibrated.

The Com-Power CDN-AF4E fully complies with the

requirements contained in the IEC 61000-4-6 and

CISPR 16-1-2.

All Com-Power CDNs can be purchased separately

or as part of the CIS series conducted immunity test

system. This is a pre-packaged solution that includes

an ACS series power amplifier and all accessories

required for the test.

Coupling Decoupling Network

CDN-AF4E

Rev. D01.14

Com-Power Corporation

19121 El Toro Road, Silverado, California 92676

(949) 459 - 9600

www.com-power.com

Advertising