C-Squared Hydrogen Discrete Analyzer Manual User Manual

Page 3

Advertising
background image

Hycision Discrete Analyzer Instruction Manual

Page 3

C-Squared Inc. 1440 S State College Blvd, Unit 4L, Anaheim, CA 92806
Tel: (714) 999-2791, Fax: (714) 999-2928,

[email protected]

,

www.c-squaredinc.com

Cautions


1. The Hydrogen Analyzer is a vibration sensitive device. Please avoid mechanical shock

while handling. Dropping the unit on a hard floor may cause damage to the unit.

2. When operating the Hydrogen unit out doors, please avoid direct sunlight on the unit.
3. The reading should be taken one minute after turning on the power switch. The power

should be turned off after taking the reading.

4. When the unit is not used, always pull up the sensor cap to allow the sensor chamber to be

exposed to ambient air.



Operating Principle



The Analyzer measures the Hydrogen content based on the extreme high thermal conductivity
of hydrogen gas. The unit consists of two main components: one is a sensor block and the
other is the electronic circuit. The patented design allows the unknown gas mixture to flow
through the sensing chamber and seal it inside for measurement. Once the gas mixture is
sealed inside the sensing chamber, the electronic circuit then measures the difference of the
thermal conductivity between the gas mixture and the reference gas. The Hydrogen content is
then calculated by the circuit and displayed.





















V:4/04/07

Advertising