Testing DC Current

1. Connect ‘Calibrated Meter’ and ‘Meter Under Test’ to the CALMASTER

right sockets

2. Switch both meter to ‘DC mA’ range
3. Confirm that the ‘Meter Under Test’ mA reading is within ±5% of the

reading displayed on the ‘Calibrated Meter’

4. Log readings obtained for calibration records
5. Repeat above on all DC current ranges on the meters

WARNING: ON 10AMP RANGE MAX TEST TIME 30 SECONDS

Fuse Protection

(AC Input) 250VAC, 1.6Amp max

(AC Output) 250VAC, 50mA max

Weight

2.2KG

Dimensions

216(L) x 132(W) x 102(D)mm

Resistance

AC Volts

DC Volts

DC Current

00.0Ω up to 100K

0.0 to 230VAC

0.0 to 36VDC

1mA
10mA
100mA
1Amp

Specifications