Trigger rules, Triggering during the idle state, Triggering during the readout state – ALLIED Vision Technologies Prosilica GT6600 User Manual
Page 71
Camera trigger
AVT Prosilica GT Technical Manual V2.1.0
71
Trigger rules
•
The end of exposure will always trigger the next readout.
•
The end of exposure must always end after the current readout.
•
The start of exposure must always correspond with the interline time if
readout is true.
•
Exposure start delay equals the readout time minus the registered expo-
sure time.
Triggering during the idle state
For applications requiring the shortest possible Trigger Latency and the smallest
possible Trigger Jitter the User Trigger signal should be applied when Imaging is
false and Idle is true. In this case, Trigger Latency and Trigger Jitter are as indi-
cated in the camera Specifications section.
Triggering during the readout state
For applications requiring the fastest triggering cycle time whereby the camera
image sensor is exposing and reading out simultaneously, apply the User Trigger
signal as soon as a valid Trigger Ready is detected. In this case, Trigger Latency
and Trigger Jitter can be up to 1 row time since Exposure must always begin on
an Interline boundary.
Trigger ready
Indicates to the user that the camera will accept
the next trigger
Registered exposure time Exposure time value currently stored in the camera
memory
Exposure start delay
Registered exposure time subtracted from the
readout time and indicates when the next exposure
cycle can begin such that the exposure will end
after the current readout
Interline time
Time between sensor row readout cycles
Imaging
High when the camera image sensor is either
exposing and/or reading out data
Idle
High if the camera image sensor is not exposing
and/or reading out data
Note
The user trigger pulse width should be at least three times the
width of the trigger latency as indicated in chapter
Term
Definition
Table 26: Explanation of signals in timing diagram