PART 4 - TROUBLE SHOOTING

CONDITION

POSSIBLE CAUSE


Unit does not turn “ON” when power is
applied

?

Verify that AC power source is live.

?

Test the fuse


FAULT CONDITION is indicated on LCD
indicator

?

Ensure that the transducer is properly
mounted to the pipe.

?

Verify that transducer connections are
correct

?

Ensure that the pipe is full of moving
liquid.

?

If SIGNAL STR is less than 100 counts
and flow rate is greater than 1 FPS [0.3
MPS], adjust GAIN control (R13 on the
Main PCB) till SIGNAL STR reaches at
least 130 counts.

?

On cleaner liquids, move the
transducers closer to a 90

o

pipe elbow.

?

If GND connection and pipe are at
different potentials, ground D77 to pipe
potential.

?

If Variable Frequency Drives are being
utilized, verify that the D77 obtains a
flow indication when the pump turns
OFF. If it does, contact the Dynasonics
factory.