Model 5861e – Brooks Instrument 5861E User Manual

Page 17

Advertising
background image

3-3

Model 5861E

Section 3 Operation

Installation and Operation Manual

X-TMF-5861E-MFM-eng
Part Number: 541B107AAG
November, 2008

To check zero always mount the controller in its final configuration and
allow a minimum of 20 minutes for the temperature of the controller and its
environment to stabilize. Using a suitable voltmeter check the controller
output signal. If it differs from the factory setting adjust it by removing the
lower pot hold plug, which is located closest to the controller body. Adjust
the zero potentiometer (refer to Figure 3-3) until the desired output signal is
obtained.

3-4 Calibration Procedure

NOTE 1: Calibration of the Model 5861E mass flowmeter requires the use
of a digital voltmeter (DVM) a flow control valve or mass flow controller to
set the flow rate and a precision flow standard calibrator such as the
Brooks Vol-U-Meter

®

. It is recommended that the calibration be performed

only by trained and qualified service personnel.

NOTE 2: If the mass flow controller is to be used on a gas other than the
calibration gas, apply the appropriate sensor conversion factor. ( Refer to
Section 4-5).

a. Adjust the anticipate potentiometer fully clockwise (20 turns). Then

adjust the anticipate potentiometer 10 turns counter clockwise to
center the potentiometer. This will provide a rough adjustment of this
circuit and make the flow signal stable for calibration.

b. Connect the DVM positive lead to the 0-5V signal output

(pin 2 D-connector, refer to Figure 3-2) and the negative lead to
circuit common (TP4). Adjust the zero potentiometer for an output of
0mV ±2mV.

c. Increase the flow rate until the flow signal output equals 5.000V.

Connect the DVM positive lead to TP2 (linearity voltage) and the
negative lead to TP4 (circuit common). Adjust the linearity
potentiometer for an output of 0.0V (zero volts).

d. Connect the DVM positive lead to the 0-5V signal output (pin 2 of D-

Connector) and the negative lead to TP4 (circuit common). Adjust
the actual flow rate to within ±2% of desired full scale flow. Measure
the flow rate using suitable volumetric calibration equipment. Adjust
the flow rate to the proper full scale flow.

Adjust the span potentiometer (Refer to Figure 3-3) until the 0-

5V

signal output equals the flow signal voltage

calculated above.

e. Connect the DVM positive lead to TP1 (sensor voltage). The voltage

at TP1 is -100 times the output voltage of the sensor. This voltage
can range from -1.2 to -12 volts, however it is recommended that
this voltage stay between -2.0 to -9.0 volts for proper operation. If
the recommended voltage range exceeds this than the desired
accuracy and/or signal stability may not be achieved. If one of the
limits is reached check the restrictor sizing procedure.
( Refer to Section 4-6).

Advertising