Section 4 Maintenance

Installation and Operation Manual

X-TMF-5850i-MFC-eng

Part Number: 541B108AAG

December, 2008

Figure 4-4 Example Nomograph

If these conditions do not exist, pressure drop equals the inlet pressure
minus the outlet pressure.

Then

Δp = 64.7 - 44.7 = 20 psi

Using the nomograph, locate the pressure drop (psi) on the vertical line
marked "

Δp" (Point A).

4. Using the nomograph, locate the pressure drop (psi) on the vertical

line marked "Dp" (Point A).

5. Locate the Nitrogen equivalent flow rate (sccm Nitrogen) on the

vertical line marked "Q

Nitrogen

" (Point B).

6. Draw a line connecting Dp and QNitrogen and extend it to the

baseline. Mark this point (Point C).

7. Locate inlet pressure (psia) on the vertical line marked "P

in

" (Point D).

8. Draw a line connection P

in

(Point D) and baseline (Point C) and then

extend this line to the vertical line marked D

o

(orifice diameter,

inches). (Point E).

9. This point on the line is the minimum orifice size for the given