Flow controllers for gas and liquid service – Brooks Instrument FC8800 Series User Manual

Page 19

Advertising
background image

2-1

Section 2 Installation

Installation and Operation Manual

X-VA-FC-eng

Part Number: 541B045AAG

March, 2012

Flow Controllers for Gas and Liquid Service

2-1 General

This section contains the procedures for the receipt and installation of the

instrument. Do not attempt to start the system until the instrument has

been permanently installed. It is extremely important that the start-up

procedures be followed in the exact sequence presented.

2-2 Receipt of Equipment

When the instrument is received, the outside packing case should be

checked for damage incurred during shipment. If the packing case is

damaged, the local carrier should be notified at once regarding his liability.

A report should be submitted to your nearest Product Service Department.

Brooks Instrument
407 W. Vine Street

P.O. Box 903

Hatfield, PA 19440 USA

Toll Free (888) 554 FLOW (3569)

Tel (215) 362 3700

Fax (215) 362 3745

E-mail: [email protected]
www.BrooksInstrument.com

Brooks Instrument

Brooks Instrument

Neonstraat 3

1-4-4 Kitasuna Koto-Ku

6718 WX Ede, Netherlands

Tokyo, 136-0073 Japan

P.O. Box 428

Tel +81 (0) 3 5633 7100

6710 BK Ede, Netherlands

Fax +81 (0) 3 5633 7101

Tel +31 (0) 318 549 300

Email: [email protected]

Fax +31 (0) 318 549 309
E-mail: [email protected]

Remove the envelope containing the packing list. Outside of your clean area,

carefully remove the equipment from the packing case. Make sure spare parts

are not discarded with the packing material. Inspect for damaged or missing

parts.

2-3 Recommended Storage Practice

If intermediate or long-term storage of equipment is required, it is

recommended that the equipment be stored in accordance with the

following:

a. Within the original shipping container.

b. Stored in a sheltered area, preferably a warm, dry, heated warehouse.

c. 32°C (90°F)maximum,45°F (7°C) minimum.

d. Relative humidity 45% nominal, 60% maximum, 25% minimum.

Upon removal from storage a visual inspection should be

conducted to verify the condition of equipment is "as received".