CHAPTER 1 - GENERAL INFORMATION

Model 12-64 Basic Flow Switch

1 - 2

Doc. No. 06EN003264 Rev. N/C

Service:

- 12-64 - Liquid, gas or slurry flow detection.
- 8-66 - Point liquid level or interface detection.

Process Connections:

1 inch male NPT.

Insertion Length:

1.2 inch (30mm) or 2.0 inch (51mm) U-length.

Material of Construction:

All wetted surfaces are 316 series stainless steel with
nickel braze per process specifications AMS 4777.

Alarm Set Points:

Setable to any value within the indicated flow range,
(for 12-64). Setable to anyinterface value, (for 8-66).
5mV hysteresis.

Time Response:

10 to 300 seconds.

Flow Rate (12-64 only):

Minimum: .010 ft/sec (.003 m/sec) in oil,
Maximum: 125 ft/sec (38 m/sec) in air.

Electrical Connection:

1 inch female NPT

Relay Rating:

DPDT contacts rated at 2A at 115 Vac.

Power Input:

100-132 Vac, 50/60 Hz, 6 watts maximum. 230 Vac
or 24 Vdc are optional power inputs.

Local Enclosure:

NEMA 4X (approximately equal to IP66) and
suitable for hazardous locations.

Operating Temperatures:

Sensing element: -100

°

to +350

°

F (-73

°

to +177

°

C).

Control circuit: -40

°

to +140

°

F (-40

°

to +60

°

C).

Operating Pressure:

To 3000 psig (207 bar).

Approvals:

FM, CSA and CE / CENELEC.

Control Circuit

The basic functions of the control circuit are to provide power to the sensing element, measure the

D

R between the

two RTDs, condition the sensing point signals, and provide relay alarm contacts for customer uses.

A double pole double throw (DPDT) relay is available in the instrument for connections to the customer alarm
systems. The relay coil can be set for either energized or de-energixed when there is either flow (wet) or no flow
(dry) of the process media. [The factory set condition is the coil being de-energized at no-flow (dry).]

The place where the relay changes state will vary depending on the type of media as well as air or liquid turbulence.
Therefore the instrument has a field adjustable alarm set point.