CHAPTER 3 - OPERATION

Model MT91 Multipoint Flowmeter

3 - 10

Doc. No. 003185 Rev. B

Ø To calibrate the analog output boards:

The analog output board calibration parameters are saved on the controller board. Therefore, analog output boards
are not interchangeable.

Tools needed:

DVM (4 1/2 digit minimum)
Calibrated 250

W

resistor

1.

Turn off the power to the flowmeter.

2.

Set both jumpers on the analog output board to the output type selected above.

3.

Connect a DVM to the appropriate connection of the analog output board. Use a precision 250ohm resistor if
the output is 4 to 20mA.

4.

Apply power to the flowmeter.

5.

Go to menu level 2.2.2 (Setup

Þ

Outputs

Þ

Analog). Select the Channel 1. Press 1 to select output signal.

6.

Select appropriate signal type. Save selection if signal type was changed.

7.

Go to menu level 4.3.1 (Calibration

Þ

CalOutputs

Þ

Analog Outputs). Select Channel 1.

8.

Press keys as indicated to reach 4mA for 4-20mA output signal or press ENTR (0) for 0-5Vdc output signal and
0-10Vdc output signal.

Note:

The DVM may read 20-30mV for 0-10Vdc output signal or 10-15mV for 0-5Vdc output signal.

9.

Press keys as indicated to reach 20mA for 4-20mA output signal or press keys as indicated to reach 5.000Vdc
for 0-5Vdc output signal or press keys as indicated to reach 10.00Vdc for 0-10Vdc output signal.

10. Repeat steps 5 through 9 for Channel 2.

Setting Discrete Outputs

Four sets of relay contacts provide the user with discrete outputs. The relays can be assigned to the parameters
marked in Table 3-3.

Table 3-3. Relay Assignment by Parameters

Relay #

Flow

Temperature Calibration Test

Zero Test

Span Test

System Status

1

2

3

4

Dependent on analog outputs

Ø To set discrete outputs

1.

Go to menu level 2.2.3.1 (Setup

Þ

Outputs

Þ

Relays

Þ

Mode)

2.

Select the relay to be set: 1=Rly #1, 2=Rly #2, 3=Rly #3, or 4=Rly #4.

Note:

Only relays 1 and 2 can be set as switch points.