Introduction

This manual guides the user of the VeriCal instrumentation through an initial gathering of in-situ baseline data. This baseline line data will then be

compared to data gathered during similar future verification processes to determine if the system is operating within factory specifications.

Theory of Operation

The VeriCal system uses a sonic nozzle to consistently control the amount of compressed air (or nitrogen) injected onto the thermal flow transducer

located on the end of the probe assembly. It is critical to use the same gas for subsequent VeriCal runs to ensure repeatability.
The operating principle of the sonic nozzle requires the total or absolute pressure on the high side of the nozzle to be greater than 20.0 PSIA. The

pressure difference between the high side of the sonic nozzle and the process pressure (low-pressure side of the nozzle) must be greater than 2:1.

When these two requirements are met, a repeatable flow is injected onto the thermal flow transducer.

Setup

FCI Recommends that this procedure be run during the commissioning process of the instrument to determine an initial installed baseline calibration

and to document any installed offset from the factory VeriCal baseline.
Frequency: Every 18 months minimum, every six months is recommended. After the process has been performed a couple times the customer should

determine the required verification frequency based upon the process conditions.
This procedure makes the assumption that the instrument has been installed and is completely functional in the normal operating condition and

orientation. The customer should also have access to the factory VeriCal calibration certificate.

Note: All standard safety procedures must be followed during the verification process. This procedure assumes the standard

packing gland process connection. Your process connections may vary. It is critical to establish a Field Baseline upon

receiving your ST100. This will ensure a greater likelihood of repeatability and establish a history of the VeriCal data.

• Apply the proper input power and allow for a 30-minute warmup. It is critical that the electronics and the sensor be fully warmed and stable

prior to the VeriCal process. Failure to allow the proper warmup time can impact repeatability.

• Loosen the packing nut on the packing gland assembly until the internal packing is lose enough to allow the probe assembly to be retracted out

of the process and is recessed completely into packing gland assembly. Retract the probe assembly completely.

Retracted VeriCal Mounting Configuration

LOCAL ENLCOSURE

VERICAL BOX

ASSEMBLY

(SHOWN WITHOUT COVER)

INTERCONNECTING

REMOTE CABLE

PACKING GLAND

ASSEMBLY

WITH 1 1/4 " NPT

REMOTE ENCLOSURE

(SYSTEM ELECTRONICS)

PLUG VALVE

(SHOWN IN

OPEN POSITION)

INTERCONNECTING,

AIR HOSE.

1/8" FEMALE NPT,

AIR INLET

BALL VALVE

PROCESS PIPING

C01061-1-1

REGULATOR

ASSEMBLY

RETRACTED

PROBE ASSY

HIGH SIDE

PRESSURE

READING

SONIC NOZZLE