Fluid Components International ST100 Series Modbus User Manual

Page 7

Advertising
background image

ST100 Series Modbus Operation

Fluid Components International LLC

3

Process Data - Table 1, cont.

Process Variable Values - Integral Registers

Variable / Parameter

Modbus Slave Register

Data Type

Access

Flow MS (VALUE)

4101

SPECIAL1

READ

Flow LS (VALUE)

4102

SPECIAL1

READ

Temperature MS (VALUE)

4103

SPECIAL1

READ

Temperature LS (VALUE)

4104

SPECIAL1

READ

Totalizer MS (VALUE)

4105

SPECIAL2 (D)

READ

Totalizer MS2 (VALUE)

4106

SPECIAL2 (D)

READ

Totalizer LS2 (VALUE)

4107

SPECIAL2 (D)

READ

Totalizer LS (VALUE)

4108

SPECIAL2 (D)

READ

Pressure MS (VALUE)

4109

SPECIAL1

READ

Pressure LS (VALUE)

4110

SPECIAL1

READ

Totalizer Value - Single Precision Floating Point (16 bits)

Variable / Parameter

Modbus Slave Register

Data Type

Access

Totalizer 1 MS (VALUE)

4111

FLOAT

READ

Totalizer 1 LS (VALUE)

4112

FLOAT

READ

Totalizer 2 MS (VALUE)

4113

FLOAT

READ

Totalizer 2 LS (VALUE)

4114

FLOAT

READ

Process Variables - Engineering Unit Codes

Variable / Parameter

Modbus Slave Register

Data Type

Access

FLOW ENG. UNITS CODE

4020

INTEGER

READ

TEMP ENG. UNITS CODE

4021

INTEGER

READ

TOTALIZER ENG. UNITS CODE

4022

INTEGER

READ

PRESSURE ENG UNITS CODE

4023

INTEGER

READ

Instrument Status Codes

Variable / Parameter

Modbus Slave Register

Data Type

Access

DEVICE (SENSOR 1) STATUS CODE #1

4025

INTEGER

READ

DEVICE (SENSOR 1) STATUS CODE #2

4026

INTEGER

READ

Note: (D) indicated double precision (64 bits)

NOTE:

DATA TYPE “SPECIAL 1”: This is a collection of discrete registers that contain a single precision (32 bit)

floating point value, and must be treated and interpreted as a single precision floating point number by

the DCS or the PLC.

DATA TYPE “SPECIAL 2”: This is a collection of discrete registers that contain a double precision (64 bit)

floating point value, and must be treated and interpreted as a double precision floating point number by

the DCS or the PLC.

Advertising