Timing, Timebase range, Combining channels – Agilent Technologies DP211 User Manual

Page 34: Trigger, Trigger source

Advertising
background image

User Manual: Family of 8-bit Digitizers

Page 34 of 66

3.3.4. DP1400 Simultaneous Multibuffer Acquisition and Readout (SMAR)

For this module the internal memory has a dual-port structure which can be exploited to permit simultaneous data
acquisition and read out. When the special SMAR option is present the memory can be turned into a circular buffer
of a chosen number of banks, between 2 and 1000. Data can be read out of one bank while data is acquired into any
available free banks. This mechanism, together with sequence acquisition, helps achieve a high maximum continuous
event rate in spite of interrupts due to the computer operating system. The maximum continuous event rate is the
maximum value of the trigger frequency that can be accepted without the loss of any event. It will work for both 1
and 2 converter/channel modes.

3.3.5. Timing

A crystal-controlled timebase is used to clock the ADC system of the digitizers. The timebase accuracy is guaranteed
as shown in the table at the beginning of this section. The digitizers also include a built-in Trigger Time Interpolator
(TTI) that measures the time from the trigger point to the first sample point. This information is essential for
determining the precise relation between the trigger or other event of interest and the digitized samples of the signal.
The approximate TTI resolution is also given in the table at the beginning of this section.

3.3.6. Timebase Range

The timebase range defines the time period over which data is being acquired. For example, the DC110 has a
standard acquisition memory of 128 Kpoints and maximum sampling rate of 1 GS/s. Therefore, at the maximum
sampling rate, the digitizer can record a signal over a timebase range of up to 130

μs (approx. 130,000 points *

1 ns/point). The timebase range can be adjusted by varying the amount of acquisition memory or the sampling rate of
the digitizer.

3.3.7. Combining channels

The DC135/DC140/DC241/DC241A/DC271/DC271A/DC271AR/DP235/DP240/DP1400 digitizers offer the
possibility of combining the converters (and their memories) from two or four channels to analyze a single input
channel. With this feature the maximum sampling rate and the maximum amount of acquisition memory can be
doubled or quadrupled if all of the input channels are not of immediate interest.

3.4.

Trigger

Normally the trigger settings applied to the digitizer are used to determine the time at which the device will stop
acquiring data. Some models are also capable of a ‘Start on Trigger’ mode of acquisition (see the Programmer’s
Guide for further details). The various trigger settings are outlined below.

3.4.1. Trigger Source

The trigger source can be a signal applied to either an Input Channel (internal triggering) or the External Trigger
Input. For the DC135/DC140/DC211/DC211A/DC241/DC241A/DC271/DC271A/DC271AR modules, a
standardized trigger in signal can also be routed via the PXI Bus Star Trigger line.

Most digitizers provide a separate front panel input BNC connector that can be used as an External Trigger Input.
The External Input provides a fully functional trigger circuit with selectable level and slope as for the Internal
Triggering source; however it does not include coupling choices nor HF, Window, and Spike Stretcher triggers. The
external trigger termination (1 M

Ω or 50 Ω) is also selectable on many modules. In modules with this feature, the

circuit also provides an overload protection that will automatically switch the coupling from 50

Ω to 1 MΩ if the

signal is greater than

±5 V DC.

The DC271-FAMILY digitizers have a fixed 50

Ω termination impedance. They also allow the same BW limiter

selections as can be found for the channels. The DC271-Family digitizers' external trigger circuit has diode
protection against overload.

The DP1400 also gives the choice of 50

Ω or 1 MΩ termination impedance for the MCX external input and has

diode protection against overload. However, there are no BW limiters and only DC trigger coupling is available.

In all 50

Ω cases a ±5 V limit on trigger signals should be respected, although somewhat higher voltages for short

time periods will not damage the unit. For 1 M

Ω input signals up to ±100 V (DC + peak AC < 10 KHz) are allowed.

Advertising
This manual is related to the following products: