Valve Issues

Problem

Potential Causes

Solutions

Valve won’t open

Air pressure too low

Ensure primary air regulator reading at least 60 psi,
turn up to 80-90psi if possible and check again

Internal valve o-ring jammed / twisted

Remove valve from manifold, Carefully remove top
of valve

(caution – under high spring pressure)

push white piston up with small allen wrench from
opposite end and check o-ring condition. Replace
and lubricate if needed.

Valve leaks air or

water out top

Internal o-ring seal damaged / worn

Remove valve from manifold, Carefully remove top
of valve

(caution – under high spring pressure)

push white piston up with small screwdriver from
opposite end and check o-ring condition. Replace
with 018 Viton O-ring and lubricate with Dow 111
valve lube.

Valve remains open

after signal is off

Manifold pressure is above 230 psi

Reduce pressure to manifold to 200 psi operating
pressure