Installation, Mechanical installation, Plumbing – Hach-Lange HACH 5500 sc SiO2 Installation User Manual

Page 7: Plumbing access ports, Removed for better access. refer to, Figure 2

Advertising
background image

Figure 2 Door removal

Installation

D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.

Mechanical installation

D A N G E R

Risk of injury or death. Make sure that the wall mounting is able to
hold 4 times the weight of the equipment.

W A R N I N G

Personal injury hazard. Instruments or components are heavy. Use
assistance to install or move.

W A R N I N G

Personal injury hazard. The object is heavy. Make sure that the
instrument is securely attached to a wall, table or floor for a safe
operation.

Install the analyzer indoors, in a non-hazardous environment. Refer to
the documentation that is supplied with the mounting hardware.

Plumbing

D A N G E R

Fire hazard. This product is not designed for use with flammable
liquids.

N O T I C E

Do not install reagents until all plumbing is complete.

Make sure to use the specified tubing size.

Plumbing access ports

Make plumbing connections through the plumbing access ports. Refer to

Figure 3

,

Figure 4

or

Figure 5

. To keep the enclosure rating, make sure

that plugs are installed in the plumbing ports that are not used.
If using an external air purge, refer to the instructions supplied with the
air purge kit to remove the fan filter and replace it with a plug. Refer to
the operations manual for instructions on enabling the air purge. Refer to
the maintenance and troubleshooting manual for the part number for the
air purge kit.

English 7