Analyzer Overview

3.11 Available options

3.11.1 K-Kit (cationic)

For a high acidity water such as that from a cation exchanger outlet, the regular gaseous
conditioning is not sufficient to raise the pH to values superior to 10.3. The forced-gas
conditioning system (K-kit) is then needed.

The K-Kit option includes a gas pump with no moving parts, a power supply board, additional
hydraulics and full installation instructions.

For further information contact your local Hach Lange representative.

3.11.2 Static heat exchanger system

A static heat exchanger system is available as an option and is capable of working with up to
four channels. It comes complete with inlet and outlet connectors (4/6mm tubing) and mounting
(2 flanges and screws).

Very easy to install, and requiring no voltage supply, this compact (350 x 40mm)
product absorbs changes of heat even on samples flowing at 5 L/h per channel.

Specially designed for POWER applications, it has a high resistance to corrosion and deposits,
and allows incoming samples from 0 - 60°C to be released to the analyzer within its operating
range of 5 - 45°C.

For further information contact your local Hach Lange representative.

3.11.3 Extra channels

The 9240 analyzer can be configured as a single or multi channel analyzer. Up to 4 channels
can be configured on a single analyzer.

Extra channels are easily installed and setup by following the instructions provided.

For further information contact your local Hach Lange representative.