Hach-Lange SI792 D Quick start guide User Manual

Quick start guide

Advertising
background image

© Hach Company, 2007. All rights reserved. Printed in Germany.

For technical assistance or ordering information: visit

www.hach.com

, call 800-227-4224 or email [email protected] Edition 1 December 2007

Quick Start Guide

For si792 D series 2-wire transmitters

DOC016.53.00797

Additional copies are available on www.hach.com

DANGER: Explosion Hazard. This Quick Start Guide is intended for setup of the si792
transmitter in a non-hazardous location. It is not a substitute for reading and following
the instructions in the user manual. Installation in a hazardous location must be
performed by qualified personnel per the applicable control drawing or ATEX certificate
in the si792 D on the provided documentation CD.

DANGER: Electric Shock Hazard. Disconnect power to all wiring before making
connection to the si792 D series transmitters.

1. Connect the power and communication (

Table 1

).

2. Connect the sensor wires (

Table 2

and

Table 3

).

3. The transmitter is set by default for Oxysens and Oxyferm. Manually select the sensor type for the
Oxygold G (si792(x) D only. Refer to the user manual for si792x D-FF and si792x D-PA.)

a. Press

CONF

,

ENTER

.

b Input passcode:

1200

. Press

ENTER

. The display will show “out.1MNU”. Press

ENTER

.

c The display will show “tyPE A” and “o1 SnSR”. Use the

ARROW

keys to select “tyPE A” (Oxy-

ferm and Oxysens) or “tyPE B” (Oxygold). Press

ENTER

.

d Press

CONF

,

ENTER

to exit the configuration mode. The transmitter will remain in hold mode

for 20 seconds, then return to measurement mode.

4. Calibrate the sensor. The two most common calibration methods are described below. Refer to
the user manual for other details.

Calibration in air (recommended)
a.
Press

CAL, enter passcode:

1100, ENTER

. The display will briefly show CAL and then the

entered (or default) humidity value.

b Remove the sensor from the process and place in air.
c Enter the actual relative humidity and press

ENTER

. The display will show the entered (or

default) barometric pressure.

d Enter the barometric pressure and press

ENTER

. A drift check will begin. The drift check may

take some time. When complete, the display will show the concentration in the units specified
in the configuration menu.

Note:To exit the drift check after more than 10 seconds, press CAL (accuracy will be reduced).

e Accept the calculated value for the concentration or adjust the value as necessary and then

press

ENTER

. The display will show the new slope and zero (relative to 25 °C and

1.013 mbar).

f Press

ENTER

to end the calibration. Return the sensor to the process. The measured value

will be shown in the display alternately with Hold.

g Press

ENTER

to return to the measure mode.

Calibration by Comparison
The sensor can be calibrated for DO (or gas volume concentration) by entering the sample
value from a measurement in an instrument such as a laboratory or portable meter. The sensor
does not need to be removed from the process during calibration. The sample temperature
should correspond to the measured process temperature for accurate results.
a. Press

CAL

, enter passcode: 1105,

ENTER

. The display will briefly show CAL PRD and then

StorE.

b Collect a grab sample for measurement in the lab. Press

ENTER

to store the sample value in

the transmitter.

c The cal mode indicator will flash to indicate that calibration is not complete. The transmitter

will continue to measure and display the current sample values.

d Measure the grab sample with a laboratory or portable meter.
e Press

CAL

, enter passcode: 1105,

ENTER

to access the product calibration once more. The

display will briefly show CAL PRD and then CALC.

f Edit the displayed value to match the value measured by the laboratory or portable meter and

press

ENTER

. The new slope and zero point will be shown.

g Press

ENTER

to end the calibration. Return the sensor to the process. The measured value

will be shown in the display alternately with Hold.

h Press

ENTER

to return to the measure mode.

Note:For comparison calibration via Foundation Fieldbus, see the user manual.

Table 1 Terminal assignments

Terminal No.

4–20 mA/HART

Profibus PA

Foundation Fieldbus

10

(–)

(–)

(–)

11

(+)

(+)

(+)

Table 2 Terminal assignments—si792(x) D (4–20 mA/HART)

Terminal No.

Oxygold G/Oxyferm sensor wire color

1

Oxysens sensor wire color

1

clear

clear

5

shield/red

brown

6

shield/yellow-green

shield

7

green

yellow

8

white

blue

1

With VP cable LZY079, LZY353 and LZY354

Table 3 Terminal assignments—si792x D-FF and si792x D-PA

Terminal No.

Oxygold G/Oxyferm sensor wire color

1

1

With VP cable LZY079, LZY353 and LZY354

Oxysens sensor wire color

2

clear

clear

4

shield/red

brown

5

shield/yellow-green

shield

7

green

yellow

8

white

blue

Advertising