Hach-Lange SI792 P series Quick Start Guide User Manual

Quick start guide

Advertising
background image

© Hach Company, 2007. All rights reserved. Printed in Germany.

For technical assistance or ordering information: visit

www.hach.com

, call 800-227-4224 or email [email protected] Edition 1 December 2007

Quick Start Guide

For si792 P series 2-wire transmitters

DOC016.53.00794

Additional copies are available on www.hach.com

DANGER: Explosion Hazard. This Quick Start Guide is intended for setup of the si792
transmitter in a non-hazardous location. It is not a substitute for reading and following
the instructions in the user manual. Installation in a hazardous location must be
performed by qualified personnel per the applicable control drawing or ATEX certificate
on the prived documentation CD.

DANGER: Electric Shock Hazard. Disconnect power to all wiring before making
connection to the si792 P series transmitters.

1. Connect the 24 VDC power and communication (

Table 1

).

2. Connect the sensor wires (

Table 2

).

Basic transmitter configuration—si792(x) P (4–20 mA/HART) only

Note: Refer to the manual for si792x P-FF (Foundation Fieldbus) and si792x P-PA (Profibus PA)

configuration instructions.

1. To enter the configuration mode, press

CONF

, configuration code:

1200

,

ENTER

. There are 4

configuration menus:

• output "out.1MNU"
• temperature compensation "tc MNU"
• calibration setup "CAL MNU"
• alarm settings "ALrt MNU"

2. In configuration mode, press the

RIGHT ARROW KEY

to navigate to a specific configuration menu.

Press

ENTER

to access the menu options.

3. Use the

ARROW KEYS

to edit values. Press

ENTER

to save the settings. If a value is outside of

the acceptable range, “Err” will be displayed and the value will not be accepted.
To exit the configuration mode and return to measurement mode, press

CONF

,

ENTER

at any time.

Configure the transmitter for ORP sensors
The transmitter has default settings for use with pH sensors. For use with an ORP sensor:
1. Select the "out.1MNU" configuration menu.
2. Change the sensor type from pH to ORP.

a. Press

CONF

,

ENTER

. Input passcode

1200

and press

ENTER

.

b Use the

RIGHT ARROW KEY

to change from pH to ORP under the "01.UniT" display code.

Press

ENTER

.

c Press

CONF

,

ENTER

to exit configuration mode.

Configure the pH buffer
For accurate pH calibration, configure the correct manufacturer of pH buffers. Calibration is rarely
performed at exactly 25 °C, and different pH buffers can exhibit different temperature dependencies.
3. Press

CONF

,

ENTER

. Input passcode

1200

and press

ENTER

.

4. Select "CAL MNU".
5. Select the appropriate code from

Table 3

.

Note:If the buffer is not listed in

Table 3

, select manual entry (MAN). Enter the pH values when

prompted during calibration.

6. Press

ENTER

. Press

CONF

,

ENTER

to exit configuration mode.

Table 1 Terminal assignments

Terminal No.

si792(x) P

si792x P-FF

si792x P-PA

14

(–)

(–)

(–)

15

(+)

(+)

(+)

Table 2 Terminal assignments—differential sensors

Terminal No. Assignment

Hach wire color

1

Measure electrode (pHD is Reference)

red

4

Reference electrode (pHD is Measure)

green

5

Cable shield connections (2 shield wires for
Hach pHD, 1 shield wire for Hach LCP)
Note: Do not connect to earth ground.

shield

8

RTD (resistive temperature device)

yellow

17

–6 V

white

18

GND

black

Table 3 Calibration mode setup menu—CAL mnu

Select pH buffer set
–01–BUF

Knick/Mettler-Toledo

–02–BUF

Merck Titrisols, Riedel Fixanals

–03–BUF

Ciba (94)

–04–BUF

NIST technical buffers

–05–BUF

NIST standard buffers

–06–BUF

HACH buffers

–07–BUF

WTW technical buffers

–08–BUF

Hamilton Duracal

MAN (manual entry, for buffers not listed in this table)
DAT—Entry of offset and slope of premeasured electrodes

Advertising