Configuration – Hach-Lange SI794 I User Manual User Manual

Page 30

Advertising
background image

30

The output current is frozen

(at its last value or at a preset fixed

value, depending on the configurati-

on), limit and alarm contacts are inac-

tive. The controller is in the configured

state, Sensoface is off, mode indicator

“Configuration” is on.

Hold

Input errors

The configuration parameters are

checked during the input. In the

case of an incorrect input “Err” is dis-

played for approx. 3 s. The incorrect

parameters cannot be stored. Input

must be repeated.

End

End with conf. The measured value

and Hold are displayed alternately,

“enter” flashes. End Hold mode

with enter . The display shows the

measured value. The output current

remains frozen for another 20 s (HOLD

icon on, “hourglass” flashes).

HOLD icon

During

configuration

the device

remains in the

Hold mode.

Configuration

In the Configuration mode you set the device parameters. The device can

store two different parameter sets and switch between them. Sensor data

and “Clean/Pset2” output are edited in parameter set 1 only. They are valid for

both parameter sets.

Press conf.

Enter mode code “1200”:

Edit parameter set 1 with and , con-

firm/proceed with enter.

Enter mode code “1288”:

Edit parameter set 2 with and , con-

firm/proceed with enter.

Configuring

Parameter set 1

Configure:

Parameter set 2

Configure:

“88” appears in the

display.