Installation

Table 3 Status register

Bit

Status 1

Description

0

Calibration activated

Calibration in progress, measurement value not up to
date

1

Cleaning activated

Cleaning in progress, measurement value not up to
date

2

Service mode activated

Device in "Service" mode, measurement value not up to
date

3

General error message

General error detected, refer to error text for details

4

Measurement value channel 0,
poor quality

Measurement accuracy is not within specified limits

5

Measurement value channel 0,
range short-fall

Measurement value falls short of the specified range

6

Measurement value channel 0,
range exceeded

Measurement value exceeds the specified range

7

Measurement value channel 1,
poor quality

Measurement accuracy is not within specified limits

8

Measurement value channel 1,
range short-fall

Measurement value falls short of the specified range

9

Measurement value channel 1,
range exceeded

Measurement value exceeds the specified range

10

Measurement value channel 2,
poor quality

Measurement accuracy is not within specified limits

11

Measurement value channel 2,
range short-fall

Measurement value falls short of the specified range

12

Measurement value channel 2,
range exceeded

Measurement value exceeds the specified range

13

Measurement value channel 3,
poor quality

Measurement accuracy is not within specified limits

14

Measurement value channel 3,
range short-fall

Measurement value falls short of the specified range

15

Measurement value channel 3,
range exceeded

Measurement value exceeds the specified range