Maintenance and Troubleshooting
42.
Rinse the sensor under a tap for about 5
seconds to remove any excess electrolyte,
then gently wipe with a soft tissue to ensure
all parts are completely dry.
43.
Drain the overflow electrolyte from the
container into a sink and flush away.
44.
Discard the used container.
45.
If not using a protection cap with grille
proceed to step
48.
Otherwise, take a new
Dacron
®
mesh from the box of O-rings in the
recharge kit.
46.
Place the mesh in the center of the
protection cap. It is very important that the
mesh is in the center of the protection cap
and covering the entire grille (as illustrated
right).
47.
Lower the sensor onto the protection cap
making sure not to disturb the mesh.