Maintenance and troubleshooting – Hach-Lange ORBISPHERE A1100 Family User Manual

Page 40

Advertising
background image

38

Maintenance and Troubleshooting

42.

Rinse the sensor under a tap for about 5
seconds to remove any excess electrolyte,
then gently wipe with a soft tissue to ensure
all parts are completely dry.

43.

Drain the overflow electrolyte from the
container into a sink and flush away.

44.

Discard the used container.

45.

If not using a protection cap with grille
proceed to step

48.

Otherwise, take a new

Dacron

®

mesh from the box of O-rings in the

recharge kit.

46.

Place the mesh in the center of the
protection cap. It is very important that the
mesh is in the center of the protection cap
and covering the entire grille (as illustrated
right).

47.

Lower the sensor onto the protection cap
making sure not to disturb the mesh.