Section 4 operation, Phosphate cuvette tests, Mode. see – Hach-Lange LCK 348_349_350 User Manual

Page 29: 2 quality assurance

Advertising
background image

29

Section 4

Operation

4.1 Analyse Phosphate-Phosphorus/Phosphate with

LANGE Cuvette Tests

DANGER
Handling chemical samples, standards and reagents can be dangerous. Users
of this product are advised to familiarize themselves with safety procedures
and the correct use of chemicals, and to carefully read all relevant material
safety data sheets.

To analyse Phosphate-Phosphorus/Phosphate (ortho or total) follow the working

procedure enclosed in the according Cuvette Test LCK 348, LCK 349 or LCK 350.

4.2 Quality assurance

The measurement results must be subjected to plausibility checks (dilute and/or

spike the sample).

Regular checks with corresponding standard solutions (see

Section 5 Replacement

Parts and Accessories on page 31

) ensure that your measurement system is

functioning properly and is being correctly operated.

4.3 Evaluate LANGE Phosphate-Phosphorus/

Phosphate Cuvette Tests

Note: Make sure that the instrument is programmed with the corresponding calibration data,
see

3.4 Check entered data on page 26

.

1.

Fill DI water into an empty cuvette (LCW 919) up to the lower edge of the

barcode label.

2.

Close the cuvette with the rubber stopper. This is the blank cuvette.

3.

Thoroughly clean the outside of the cuvette.

4.

Press POWER to switch on the instrument.

5.

Make sure that the correct measuring range is selected!

Otherwise press MENU. "SEL" is displayed. A flashing arrow indicates the

current measuring range (1 or 2). Press GREEN key to select measuring range.

Press MENU to get back to the measuring mode.

6.

Insert the blank cuvette into the cell holder equipped with the cuvette adapter

1 cm.

Note: Make sure that the cuvette adapter is positioned correctly.

7.

Place the instrument cap with the HACH LANGE label over the cell holder to

cover the cuvette.

8.

Press BLUE key. "0.00" is displayed.

9.

Remove the instrument cap and insert the sample cuvette.

10. Place the instrument cap over the cell holder to cover the cuvette.
11. Press GREEN key.

Advertising