Inputs/Outputs Menu

The following table lists the default configuration. The first three events on the list are pre-set
and only the priority can be changed:

Analog outputs (continued)

Channel configuration

Set the type of measurement that will be transmitted through each output
channel, and the output characteristics.

Meas. type

: Select between the type of measurements available in the

rolling list.

Characteristics

: Select either Linear, Tri-linear or None (see

Analog

output characteristics on page 61

).

Limits

: Press the configure button to adjust the analog output set points

for each output. Enter values in the appropriate text boxes. In Linear
mode, only the ML and MH values can be adjusted. Tri linear mode
allows all limits to be adjusted, and the None mode denies access to this
screen.

The authorized user may define a maximum of 12 customized events for
each analog output and change the order of priority of all events.
Note:

This only applies to Tri linear and Linear outputs. It is not available if

the output characteristic is set to None.
Configure the events that should be signaled at the corresponding current
shown in the right column.

Only one event signal at a time can be sent via the current output. As
there is a possibility to have several events at the same time, an order of
priority must be set. This order has been set by default, but it can be
modified to suit particular needs and conditions. Touch the priority
number in the left column and edit it.

The shaded events in the list have preset outputs and only the priority
can be changed. The other events can be customized by the user.
Touch a white text box to call up the rolling list. Select an event from this
list and press OK. Then adjust the priority as required.

Note:

When an event occurs, measurement information is superseded by

the event information on the output.

Table 2 Extended event table

Priority

Event

I [mA]

0

Sensor out

0.25

1

External pressure sensor out

0.50

2

Interfering gas error

0.75

3

Custom Event 1

1.00

4

Custom Event 2

1.25

5

Custom Event 3

1.50

6

Custom Event 4

1.75

7

Custom Event 5

2.00

8

Custom Event 6

2.25

9

Custom Event 7

2.50

10

Custom Event 8

2.75

11

Custom Event 9

3.00

12

Custom Event 10

3.25

13

Custom Event 11

3.50

14

Custom Event 12

3.75