2 calibration in line, Calibrations 6.4.2 calibration in line – Hach-Lange ORBISPHERE 3650 User Manual

Page 45

Advertising
background image

43

Calibrations

6.4.2 Calibration in line

The in line calibration procedure can be used to calibrate the sensor directly in line, against a
liquid sample with a known dissolved gas concentration. To perform this type of calibration, the
instrument must be set for calibration In line. Switch on the instrument, if necessary, and wait a
minute or so for the displayed measurement to settle. Expose the sensor to a liquid sample with
a known gas concentration.

1.

Press the

CAL

button. Remember, this

button may have been locked out to prevent
an accidental reset (see

Locking out the

instrument’s CAL button on page 39

for

details).

2.

A brief clearing [---] message appears.

3.

Press

CAL

again within a 3 second period.

4.

A measurement will flash on the LCD,
showing the gas concentration of the
calibration sample.

5.

Assuming you know the gas content to be a
certain value, you can adjust the displayed
value with the

Up/Down Arrow

keys.

6.

When the reading is adjusted to the known
concentration, press

CAL

again.

7.

The instrument displays CAL and returns to
the measurement mode.