2 calibration in line, Calibrations 6.4.2 calibration in line – Hach-Lange ORBISPHERE 3650 User Manual
Page 45
43
Calibrations
6.4.2 Calibration in line
The in line calibration procedure can be used to calibrate the sensor directly in line, against a
liquid sample with a known dissolved gas concentration. To perform this type of calibration, the
instrument must be set for calibration In line. Switch on the instrument, if necessary, and wait a
minute or so for the displayed measurement to settle. Expose the sensor to a liquid sample with
a known gas concentration.
1.
Press the
CAL
button. Remember, this
button may have been locked out to prevent
an accidental reset (see
instrument’s CAL button on page 39
details).
2.
A brief clearing [---] message appears.
3.
Press
CAL
again within a 3 second period.
4.
A measurement will flash on the LCD,
showing the gas concentration of the
calibration sample.
5.
Assuming you know the gas content to be a
certain value, you can adjust the displayed
value with the
Up/Down Arrow
keys.
6.
When the reading is adjusted to the known
concentration, press
CAL
again.
7.
The instrument displays CAL and returns to
the measurement mode.