Section 1

Specifications

1.1 Technical data on the 1200-S sc - pH / ORP sensor

Specifications are subject to change without notice.

Materials

Stainless steel metal housing

Enclosure rating

IP 68; stainless steel metal housing

pH measuring range

0 pH ... 14 pH

ORP measuring range

–1,500 ... 1,500 mV

Temp. measuring range

–5 °C ... 50 °C

Storage temperature
Sensor and controller

–20 °C ... 60 °C; 95 % relative humidity, non-condensing

pH / ORP response time

< 15 s; T90

Temp. response time

< 2 min; T90

pH measuring accuracy

± 0.02 pH

ORP measuring accuracy

± 1.2 mV

Temp. measuring accuracy

± 0.2 °C

Reproducibility

± 0.5 % of the end value of the measuring range

Sensitivity

± 0.5 % of the end value of the measuring range

Calibration, pH

One or two point, standard buffer solution (automatic),
One or two point, comparative measurement

Calibration, ORP

One point, comparative measurement

Calibration, temp

One point, comparative measurement

Max. immersion depth / pressure
for the sensor

20 m / 2 bar over-pressure

Maximum flow speed

4 m/s

Sensor interface

MODBUS

Sensor cable

10 m, hard wired, polyurethane

Sensor weight

<1 kg

Sensor dimensions (Ш Ч L)

42 × 504 mm

Fastening

• Immersed pipe
• Chain

Service life measuring electrode

Approx. 1 year

Sensor power

< 7 W