Observation

Explanation

The bottle goes up with the piercing head

Stop the analysis by pressing the emergency STOP
button.
Disconnect instrument from the power supply and
remove the bottle. Check that the piercing spike is in
place and not unscrewed. Release emergency STOP
button by turning it clockwise and restart the
instrument.

Open the front door to stop the process. Remove the
bottle and check that the piercing spike is in place and
is not unscrewed. Close the front door and press the
Abort button on the screen.

No antifoam being injected

Go to the antifoam section of the maintenance wizard
and prime the pumps until antifoam can be seen being
regularly injected into the antifoam priming beaker.