Probe calibration

Whenever possible, we advise proceeding in
the following order:

Temperature calibration

¾ It's an indispensable operation during

the commission for taking into account
the resistivity of the cable and offset
probe Pt100.

¾ Immerse the probe in a solution for

about 10 mn.

¾ Record the temperature of the solution

with a thermometer (precision < ± 0.1°C).

¾ Programm the transmitter in process

calibration mode.

¾ Adjust the value of the temperature read

with that of the thermometer.

Conductivity calibration

First method (recommended) :

¾ Programm the transmitter in electrical

calibration mode. Choose the resistance
the closest possible to your process
(see table on the following page).

¾ First point: Remove the probe from the

liquid or unscrew the connector (Infinite
resistance, taking into account the
capacity of the cable).

¾ Second point: Connect the resistance

(precision <

0,1

%) of the value

programmed at the IN/OUT terminals of
the conductivity module.