Section 1

Specifications

Specifications are subject to change without notice.

Components

Corrosion-resistant materials, fully-immersible probe with 10 m (30 ft) cable

Measuring Range (Dissolved Oxygen) 0 to 40 ppm (0 to 40 mg/L) or 200 % saturation

Measuring Range (Temperature)

– 5 to 50 °C (23–120 °F)

Probe Operating Temperature

– 5 to 50 °C (23–120 °F)

Probe Storage Temperature

– 5 to 70 °C (23–158 °F); 95 % relative humidity, non-condensing

Response Time at 20 °C

120 seconds to 90 % of value upon step change

Measurement Accuracy

± 2 % of span

Temperature Accuracy

± 0.2 °C

Temperature Compensator

30K NTC Thermistor

Repeatability

± 0.5 % of span

Sensitivity

± 0.5 % of span

Calibration

Air/Sample

Maximum Pressure

10 bar (145 psi)

Probe Cable Length

Integral 33 ft (10 m)

Probe Weight

0.26 kg (9.1 oz)

Probe Dimensions

See

Figure 3 Probe Dimensions on page 8

.

Wetted Materials

Noryl, PVC, Viton, Polypropylene, Nylon

Electrode Materials

Nickel-chrome and lead

Minimum Flow Rate

0.5 cm/s (0.016 ft/s)

Measuring Principle

Galvanic